
DG Spin
Procedure for Qualification of the Operation (OQ)
8-4
•
Verify that centrifugation is stopped and that the out-of-balance
indicator (
) is activated in the Warning area of the Display during
centrifugation
NOTE:
The centrifugation stop enables ensuring
correct function of the
DG Spin
centrifuge with
respect to out-of-balance control system.
Otherwise, contact your authorised Technical
Service.