DG Spin
Installation
5-2
5.3
Unpacking the equipment
Once the appropriate place is selected for installation, unpack the equipment
according to the following instructions:
•
Open the upper part of the box.
•
Remove all protections and accessories.
•
Remove the protection bag and place the equipment in the chosen
position.
•
Verify that all elements specified in the attached
Packing list
are
included.
•
Check that the equipment is not damaged due to the storage and
transport.
5.4
Installation procedure
To install the
DG Spin
centrifuge, proceed as follows:
•
Connects the mains cord (Ref. (3), Figure 2) to the centrifuge and the
mains.
•
Switch on the Main switch (Ref. (6), Figure 2).
•
Verify that the Display shows the centrifugation time.
•
Press
and remove the internal protection that protects the
DG Spin
Head.
•
Remove the
DG Spin
Head from the Centrifugation chamber and
verify that all the Card holders (Ref. (2), Figure 4) of the
DG Spin
Head
are perfectly placed and held by the corresponding pivot.
•
Close the Centrifuge cover by pressing the front central part.
•
Ensure that the Cover is completely closed and press the Centrifugation
starting button (
).
•
Verify that the
DG Spin
Head starts to rotate.
•
After 5 minutes, press
again and verify that the centrifugation
speed that appears in the Display is within the speed range defined in
3.1"Technical specifications".
•
Later on, stop centrifugation by keeping
pressed for a few seconds.
The centrifuge will emit some intermittent acoustic signals that will end
in a continuous signal that indicates the centrifugation stop.