DG Spin
Procedure for Qualification of the Operation (OQ)
8-1
8
OPERATION QUALIFICATION PROCEDURE (OQ)
The
DG Spin
centrifuge has two non-adjustable independent time basis controlled
by the centrifuges program itself. Given that the speed has been calibrated by the
manufacturer, if no function alarm is activated (see sections 12.1 and 12.2), it is
warranted that
DG Spin
centrifuge will correctly work for all its useful life, without
any additional calibration.
However, the procedure described below aims to provide a method for those
laboratories which, under a quality management system, should periodically verify
that the
DG Spin
centrifuge works under specifications.
The Operation Qualification Procedure (OQ)
of the
DG Spin
centrifuge consists
of verifying the correct function of the following parameters:
-
Centrifugation speed
-
Balance control system
8.1
Required materials
To qualify operation of the
DG Spin
centrifuge the following materials are
required:
-
Calibrated optical tachometer
-
24
DG Gel
cards
To warranty the veracity of measurements, these materials should be calibrated
using instruments controlled regularly with traceable materials at primary or
secondary reference standards.
NOTE:
The use of any tachometer is possible as long
as it complies with the following specifications:
Tachometer: Optical
Scale division (rpm):
≤
1 rpm
Range (rpm): 1.00 - 99,999 rpm
Maximum uncertainty permitted:
± 3 rpm
Detection distance: < 35 cm with visible red
light