background image

Page  3

OUTPUT SCALING

The CO

2

 output signal is scaled such that 4-20 mA (or 0-5 / 0-10 Vdc) equals 0-2000 ppm by default. The 

signal span may be changed via the Setup Menu from 1000 to 10,000 ppm in 500 ppm increments.

START-UP MODE

Verify the device is properly wired and connections are tight. Ensure the mA / VOLT switch is set for the 

correct signal type. Apply power and the LCD will indicate the software version number, the output signal 

type and then the sensor will begin a one minute warm-up countdown. When the warm-up time expires 

the device will enter normal mode.

NORMAL MODE

The sensor will begin reading the CO

2

 level (and temperature if it is installed), output the correct analog 

signal and display the values on the LCD. If the relay is installed, the measurements will be compared to 

the setpoint and appropriate action taken. The keypad is monitored and if the <MENU> key is pressed the 

device will enter the Setup Menu. The Setup Menu can be accessed to configure the device for required 

operation.

RELAY MODE

The optional relay can be used as an alarm signal to indicate that a setpoint has been exceeded. For 

the basic CO

2

 product, the relay may be configured via the Setup Menu as a high or low alarm and the 

setpoint value, hysteresis and on-delay time may be set. The relay will activate and close the contacts if 

the CO

2

 measurement goes above the setpoint (for a high alarm) or below the setpoint (for a low alarm). 

If required, this action can be delayed by a time period (relay delay time) to prevent false alarms. After the 

relay is activated, the measurement must fall below the setpoint - hysteresis value (for a high alarm) or go 

above the se hysteresis value (for a low alarm) to deactivate the relay and open the contacts. The 

high/low assignment, setpoint, delay time and hysteresis values may all be programmed via the Setup 

Menu as required.

If the optional temperature sensor is installed, the relay operation may be assigned to operate as a 

temperature high or low alarm instead of a CO

2

 alarm. All the same relay parameters may be configured 

to set up the alarm as required.

SETUP MENU

The device has several parameters that can be configured locally via the Setup menu using the keypad 

and LCD. Any changes made are saved in non-volatile memory and are restored in case of a power loss. The 

menu can be accessed at any time after the start-up mode and if there is 5 minutes of inactivity the menu 

will close and normal operation will continue.

To enter the menu, press and release the <MENU> key. This will enter the Setup Menu step 1, pressing the 

<MENU> key a second time saves the setting and advances to step 2. Each press of the <MENU> key saves 

the current setting and advances the menu item. The <UP> and <DOWN> keys are used to make changes 

to program variables by scrolling through the available options. The first column below shows what will 

be displayed on the LCD, including the default value.

<MENU> 

Press and release the <MENU> key to enter the Setup Menu

CO2 Span

2000 ppm

Self-Cal

ON

The CO

2

 Span value controls the analog output scaling. CO

2

 span can be changed

from 1000 to 10,000 ppm in 500 ppm increments using <UP> or <DOWN>. If CO

2

 

Span = 2000 ppm, then the output signal is scaled 0-2000 ppm = 4-20 mA. Factory 

default is 2000 ppm.

The Self-Cal feature <MENU> when enabled requires the sensor to be exposed 

to fresh air for at least one hour every twenty-four hours. When first installed the 

sensor requires a minimum of seven days of operation with fresh air exposer daily 

to allow the algorithm to adjust for self-calibration. Default for device is having this 

feature disabled enabled “ON". If disabling this feature a manual calibration should 

be performed to ensure accuracy of the device.

<MENU> 

Press to advance to next menu item

<MENU> 

Press to advance to next menu item

1. SPAN

2. SELF-CALIBRATION

The optional relay output connects to the RELAY terminals. The relay output is completely isolated and 

has a Normally Open (NO) signal. This signal can be used to directly control an alarm or ventilation fan. See 

specifications for contact ratings. See Figure 8.

The optional two-wire temperature sensor output is available with various RTDs and thermistors to suit all 

control applications and is available on the TEMP terminals. See Figure 9.

Copyright © Greystone Energy Systems, Inc.   All Rights Reserved    Phone:  +1 506 853 3057  Web: www.greystoneenergy.com

IN-GE-CD2DTXXX-03

Summary of Contents for CD2DT Series

Page 1: ... intended for life safety applications NOTE This CO2 sensor incorporates a Self Calibration feature to correct CO2 sensor drift This feature is recommended for applications where the CO2 will be exposed to fresh air 400 ppm at least one hour per day If the monitored space is occupied 24 hours or consistently maintains higher or lower levels of CO2 it is recommended that this feature be turned off ...

Page 2: ...ram See Figure 5 This is a 3 wire sourcing device Connect the plus DC or the AC voltage hot side to the PWR terminal and the common is connected to the COM terminal The device is reverse voltage protected and will not operate if connected backwards It has a half wave power supply so the supply common is the same as the signal common Several devices may be connected to one power supply and the outp...

Page 3: ...ired SETUP MENU The device has several parameters that can be configured locally via the Setup menu using the keypad and LCD Any changes made are saved in non volatile memory and are restored in case of a power loss The menu can be accessed at any time after the start up mode and if there is 5 minutes of inactivity the menu will close and normal operation will continue To enter the menu press and ...

Page 4: ...actual output signal will depend on the PCB switch position mA or VOLT and the V Output setting for voltage operation Press MENU to set it back to OFF and advance Factory default is OFF MENU Press to advance to next menu item MENU Press to advance to next menu item 6 OUTPUT TEST Temp Units C Display CO2 ppm CO2 1235 ppm Temp 24 3 C 1235 ppm 24 3 C CO2 1235 ppm Temp 24 3 C The Temp Units setting is...

Page 5: ...nge the setting within the limits as shown in the table Note that only the one previous selection will be available Note that a high alarm activates when the measurement exceeds the setpoint and a low alarm activates when the measurement goes below the setpoint Use MENU to save and advance or 12 RELAY SETPOINT RELAY ASSIGNMENT DEFAULT SETPOINT SETPOINT RANGE RESOLUTION Temp High 26 0 C 79 0 F 5 0 ...

Page 6: ...ue Another method is to determine the CO2 ambient level using a calibrated unit as a reference and adjust calibration setting to desired value MENU Press to advance to exit the Setup Menu and return to normal mode 17 CALIBRATE 16 CALIBRATE CALIBRATION Calibration with gas requires a field calibration kit with pressure regulator tubing sensor adapter and bottle of CO2 gas The certified calibration ...

Page 7: ...ve 24 Vdc Current 550Ω max Voltage 10 000Ω min Ambient Operating Range 0 to 50 C 32 to 122 F 0 to 95 RH non condensing Storage Conditions 40 to 70 C 40 to 158 F 0 to 95 RH non condensing LCD Display Units ppm CO2 C F for optional temperature setpoint Display Range 0 10 000 ppm 0 to 50 C 0 to 122 F Display Size 35mm W x 15mm H 1 4 x 0 6 Digit Height 2 line x 8 character OPTIONAL TEMPERATURE SENSOR ...

Reviews: