5.2 Min-/Max-Alarm
This chapter describes how t
o configure the device„s alarm boundaries for min-/max-alarm-monitoring.
This instruction demands that you selected “
AL
“ as your desired output function.
- Press button 1
(when not already done)
, the device will displa
y “
AL.Hi
“. (maximum alarm-value)
- Use button 2 and button 3 to set the desired value, the device should turn on its maximum-alarm.
-
Press button 1 to validate your selection. The display shows “
AL.Hi
“ again.
- When pressing button 1 again, the device wil
l display “
AL.Lo
“. (minimum alarm-value)
- Use button 2 and button 3 to set the desired value, the device should turn on its minimum-alarm
-
Press button 1 to validate your selection. The display shows “
AL.Lo
“ again.
- When pressing button 1 again, the dev
ice will display “
A.dEL
“. (delay of the alarm-function)
- Use button 2 and button 3 to set the desired delay of the alarm-function.
Note:
the selected value [0 ... 9999] accords the alarm delay in seconds.
The device will turn on the alarm after minimum or maximum alarm value was active for the
delay-time you have set.
-
Press button 1 to validate the delay time. The display shows “
A.dEL
“ again.
Example: You want to have a temperature alarm-monitoring of a greenhouse. The alarm should start when
the temperature rises above 50°C or falls below 15°C.
Therefore your settings will be 50°C for the maximum alarm-
value “AL.HI“ and 15°C for the mini-
mum alarm-
value “AL.Lo“.
The alarm will be starting after the temperature rises above 50°C and stays above 50°C for the
entered delay time or after it had been falling below 15°C and stays below 15°C for the entered
delay time.
Please note that the alarm-outputs are inverted! This means, that the output will be active when there is no
alarm!
Now you finished configuring the alarm adjustment of your device. Press button 1 to finish the adjustment
and to switch over to display the measuring value.