8 Specification

Measuring ranges:

Display range:

max. -1999...9999 digit, depending on connected sensor

Resolution:

depending on connected sensor

Pressure units:

mbar, bar, kPa, MPa, mmHg, PSI mH

2

O (only for devices with ‚m‘ printed below

display), selectable depending on connected sensor

Accuracy: (typ.)

±0,1%FS  

(at nominal temperature)

(FASt and P.dET:  ±0.5%FS)

Measuring rate:

4 meas./sec

Nominal temperature:

25°C

Sensor:

All sensors of the GMSD..-series without recalibration can be connected

Connection: 

Mini-DIN-Socket with locking mechanism
The sensor will automatically be detected, the measurement range settings are set
referring to sensor data

Power-Off-Function:

  Device will be automatically switched off if no key is pressed/no interface

communication takes place for the time of the power-off delay. The power-off delay
can be set to values between 1 and 120 min.; it can be completely deactivated.

Display

:

2 four digit LCDs (12.4mm high and 7 mm high) for measuring values, and for min/
max memories, hold function, etc. as well as additional functional arrows.

Pushbuttons

6 membrane keys

Interface

:

Serial interface (3.5mm jack) can be connected to RS232 interface of a PC via
electrically isolated interface adapter GRS3100, GRS3105 or USB3100 (see
accessories).

Power supply

9V battery, type: IEC 6F22  (included in scope of supply)
as well as additional d.c. connector (diameter of internal pin 1.9 mm) for external
10.5-12V direct voltage supply.               (suitable power supply: GNG10/3000)

Power consumption: 

Slow measuring rate: < 1.5 mA

Low battery warning:

!

-display and ' bAt '

Housing

impact-resistant ABS, membrane keyboard, transparent panel, Front side IP65

Dimensions: 

142 x 71 x 26 mm (L x W x D)

Working temperature:

-25...+50°C

Allowable rel. humidity: 0...95 % RH (not condensing)
Storage temperature:

-25...+70°C

EMC

:

The device corresponds to the essential protection ratings established in the
Regulations of the Council for the Approximation of Legislation for the member
countries regarding electromagnetic compatibility (2004/108/EG).

Additional fault: 

<1%

_

+

Summary of Contents for GMH3110

Page 1: ...H31 0 11 6C 11 User s Manual Handheld Digital Pressure Meter GMH3110 Version 5 1 for GMSD Pressure Sensors...

Page 2: ...densation may cause in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to...

Page 3: ...he left side of the device 1 4 Display 1 5 Basic Operation When switching on the device and a zero point adjustment was carried out it shows shortly nuLL Corr On Off Switch Tara Calling of tara functi...

Page 4: ...P oFF Auto Power Off Time The device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off time The power off time can be set to...

Page 5: ...ce and a suitable electrically isolated interface adapter GRS3100 GRS3105 or USB3100 the device can be connected to a computer for data transfer With the GRS3105 up to 5 devices of the GMH3xxx series...

Page 6: ...sured value far out of allowable range Check pressure not within sensor range Battery empty Replace battery Mains operation without battery wrong voltage or polarity Check power supply replace it when...

Page 7: ...o values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7 mm high for measuring values and for min max memories hold function etc as well as additiona...

Page 8: ...BAE 0 0 400 0 bar abs 0 1 bar 600 bar abs C Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1...

Reviews: