GREISINGER GMH3110 User Manual Download Page 8

H31.0.11.6C-11

User’s Manual GMH3110

page 8

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

  

Sensors  

(03/2005)

Type

Measuring Range

Resolution

Overload

Description

GMSD 2.5 MR

-1.999 ... +2.500 mbar rel.

0.001 mbar

250 mbar rel.

A

GMSD 25 MR

-19.99 ... +25.00 mbar rel.

0.01 mbar

350 mbar rel.

A

GMSD 350 MR

-199.9 ... +350.0 mbar rel.

0.1 mbar

1 bar rel.

A

GMSD 2 BR

-1.000 ... +2.000 bar rel.

1 mbar

4 bar rel.

A

GMSD 10 BR

-1.00 ... +10.00 bar rel.

10 mbar

13.5 bar rel.

A

GMSD 1.3 BA

0 ... 1300 mbar abs.

1 mbar

4 bar abs.

A

GMSD 2 BA

0 ... 2000 mbar abs.

1 mbar

4 bar abs.

A

GMSD 7 BA

0.00 ... 7.00 bar abs.

10 mbar

10 bar abs.

A

GMSD 350 MRE 0.0 ... 350.0 mbar rel.

0.1 mbar

1.3 bar rel.

B

GMSD 3.5 BRE

0 ... 3500 mbar rel.

1 mbar

7 bar abs.

B

GMSD 1 BAE

0 ... 1000 mbar abs.

1 mbar

2 bar abs.

B

GMSD 3.5 BAE

0 ... 3500 mbar abs.

1 mbar

7 bar abs.

B

GMSD 7 BAE

0 ... 7000 mbar abs.

1 mbar

13.5 bar abs.

B

GMSD 35 BAE

0.00 ... 35.00 bar abs.

10 mbar

58 bar abs.

C

GMSD 70 BAE

0.00 ... 70.00 bar abs.

10 mbar

100 bar abs.

C

GMSD 160 BAE 0.0 ... 160.0 bar abs.

0.1 bar

600 bar. abs.

C

GMSD 250 BAE 0.0 ... 250.0 bar abs.

0.1 bar

600 bar. abs.

C

GMSD 400 BAE 0.0 ... 400.0 bar abs.

0.1 bar

600 bar. abs.

C

Description A:

Sensor is suitable for air, non corrosive and non ionising gases and liquids
Pressure connection:  2 nylon pressure ports for connection to 6 x 1 mm tubes
Housing ABS with mounting eyelet, dimensions 68 x 32.5 x 27.5 mm

Description B:

Stainless steel sensor. Suitable for aggressive media, water, etc.
Pressure connection: Threading G¼", for open ended spanner size: 27 mm
Housing ABS, dimensions ca. Ø26 (31) x 103 mm

Description C:

Stainless steel sensor. Suitable for aggressive media, water, etc.
Pressure connection: Threading G¼", for open ended spanner size: 27mm
Steel housing, dimensions ca. Ø26 (31) x 110 mm

10     Accessories

GKK3000

 Suitcase 

(275 x 229 x 83 mm)

 with punched lining suitable for the GMH3xxx-series.

GKK3100

 Suitcase 

(275 x 229 x 83 mm)

 with foam lining for universal applications.

GMH1300

Magnet holder

GAK9V

 

Rechargeable accu 9V

GLG1300

 

Accu charger for recharging 2 accus at one time

GNG10/3000

 

power supply 10V/10mA

GRS3100

 

Interface converter, electrically isolated

GRS3105

 

5-way interface converter, electrically isolated

GAM3000

 

Control device for GMH3xxx-devices with alarm function

EBS9M

9-channel software to display the measuring values

GMH3000.DLL

universally applicable windows functions library ('GMH3000.DLL') with documentation for
your own software applications

Summary of Contents for GMH3110

Page 1: ...H31 0 11 6C 11 User s Manual Handheld Digital Pressure Meter GMH3110 Version 5 1 for GMSD Pressure Sensors...

Page 2: ...densation may cause in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to...

Page 3: ...he left side of the device 1 4 Display 1 5 Basic Operation When switching on the device and a zero point adjustment was carried out it shows shortly nuLL Corr On Off Switch Tara Calling of tara functi...

Page 4: ...P oFF Auto Power Off Time The device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off time The power off time can be set to...

Page 5: ...ce and a suitable electrically isolated interface adapter GRS3100 GRS3105 or USB3100 the device can be connected to a computer for data transfer With the GRS3105 up to 5 devices of the GMH3xxx series...

Page 6: ...sured value far out of allowable range Check pressure not within sensor range Battery empty Replace battery Mains operation without battery wrong voltage or polarity Check power supply replace it when...

Page 7: ...o values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7 mm high for measuring values and for min max memories hold function etc as well as additiona...

Page 8: ...BAE 0 0 400 0 bar abs 0 1 bar 600 bar abs C Description A Sensor is suitable for air non corrosive and non ionising gases and liquids Pressure connection 2 nylon pressure ports for connection to 6 x 1...

Reviews: