Page 12 of 16

 

 

How to carry out calibration 

Please note: the calibration is only possible, if the logger memory is empty. 

Start calibration: press "CAL" (key 3) for 2 sec. 

(after more than 10 sec. the factory calibration will be set) 

The display prompts you to measure the first humidity value. Use "Set"-key (key 4) to stop calibration whenever you 
want to. In such a case the last calibration before this one will be used.

 

 

1) Selection automatic detection / manual input 

Press "CAL"-key (key 3) for a short time to switch over between the various possibilities existing: 

 

automatic detection 

(acceptable humidity variables see above) 

Display will switch over between the acceptable variables. 

 

manual input 

 

If you want to use other humidity values than those provided in the automatic detec-
tion, please enter them here. 
 

0 ... 100.0 %: input range for rel. atmospheric humidity. 
(please note: Watch out for  'Automatic temperature compensation during 
calibration') 

 

2) Calibration point 1

 

 

Put sensor in suitable calibration equipment

 

As long as the individual values in the display for the automatic detection keep changing, a 
valid value could not be detected yet (humidity value measured may deviate from value set 
by manufacturer by approx. 10%)

 

In case of manual input, enter value here

As soon as the display stops blinking and changing between values, a stable value has been de-
tected and can be taken over by means of the "Store"-key (key 6). Then the next calibration step will 
be displayed. 

3) Calibration point 2

 

 

Put sensor into suitable calibration equipment prepared for the second humidity value. 
Precondition: If the first value was below 50%, this value has to be over 50% or vice versa.  
Except this, same procedure as above. As soon as the display stops blinking and chang-
ing between values, the measuring value can be taken over by means of the "Store"-key 
(key 6) and the calibration has been completed. 

If error messages are displayed when calibrating the instrument, the old calibration keeps valid, the new calibration 
data are lost. Please refer to "Error and System Messages During TFS0100 Calibration" in chapter 12. 
 

10  System And Error Messages 

10.1  Messages at device startup 

Message (display) 

Description

 

segment test  

(8888 and all special sign’s/arrows) 

 

identified sensor  

(tFS 0100, StS 005 or StS020) 

 

temperature offset of the TFS 

(display see chapter 5)  

 

only with TFS0100 and adjusted offset-value <> off 

flow - averaging procedure 

(AVG Hold or AVG Cont)  

only with STS... 

flow - averaging period 

(display see chapter 5)

 

only with STS... 

temperature offset for NiCr-Ni-probe

 

(display see chapter 5)

 

only at adjusted offset-value <> off 

display correction for NiCr-Ni-probe 

(display see chapter 5)

 

only at adjusted corr-value <> off 

 

 

r.H.

Td

T2

Td

m/s

kJ/kg

AL

Logg

T1

Corr

CAL

r.H.

Td

T2

Td

m/s

kJ/kg

AL

Logg

T1

Corr

CAL

max

min

2

5

%

r.H.

Td

T2

Td

m/s

kJ/kg

AL

Logg

T1

Corr

CAL

%

r.H.

Td

T2

Td

m/s

kJ/kg

AL

Logg

T1

Corr

CAL

Summary of Contents for GMH 3330

Page 1: ...93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Digital Precision Hand held Measuring Device For Atmospheric Humidity Temperature Dew Point Dew Point Distance Enthalpy and Flow...

Page 2: ...COMBINATION MEASURING SENSOR TFS0100 9 7 MEASUREMENTS USING FLOW MEASURING PROBES STS005 STS020 10 8 NOTES TO SPECIAL FUNCTIONS 11 8 1 ZERO DISPLACEMENT OFFSET 11 8 2 DISPLAY CORRECTION FACTOR CORR 1...

Page 3: ...ombined with the combined measuring probe TFS 0100 is designed for room climate meas urements This includes gauging of atmospheric humidity temperature dew point dew point distance and enthalpy Togeth...

Page 4: ...n lead to malfunctions or destroying of the device and the connected devices This device must not be run with a defective or damaged power supply unit Danger to life due to electrical shock 4 Do not u...

Page 5: ...g 4 1 Display Depending on the measuring probes sensors connected the following measuring results can be displayed TFS 0100 Main display r H relative atmospheric humidity in Secondary display possible...

Page 6: ...ing press Menu for 2 sec configuration menu is activated Store Quit Measurement Hold current measuring value HLD in display for flow meas in AVGHold mode start new measurement or handling of logger fu...

Page 7: ...tc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 Device attached to...

Page 8: ...ll be multiplied by this factor off Factor is deactivated 1 000 Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time h...

Page 9: ...ergy con tent of air with a relative humidity of 0 and 0 C is 0kJ kg The warmer the air the higher the relative hu midity the higher the energy content Therefore more energy is required to heat up hum...

Page 10: ...tions tend to be quite high To be able to display a stable measur ing value two averaging functions have been integrated in the instrument Continuous Averaging The average value displayed has been cal...

Page 11: ...sses of all devices must not be identical In case several devices will be connected via one interface make sure to configurate the base addresses accordingly Channel 1 will be addressed by the base ad...

Page 12: ...values a stable value has been de tected and can be taken over by means of the Store key key 6 Then the next calibration step will be displayed 3 Calibration point 2 Put sensor into suitable calibrat...

Page 13: ...cable defective replace Values below measuring range check are there any values below the measuring range specified meas value too low Sensor cable defective replace System fault switch on again if fa...

Page 14: ...Windows 7 Windows 8 8 1 Windows 10 Programming examples Visual Basic 4 0 Delphi 1 0 The following interface functions will be supported Channel Dll Code Name function 1 2 3 4 5 6 x x x x x x 0 Read n...

Page 15: ...lip for table top or sus pended use Weight approx 160 g Interface serial interface 3 5mm jack serial interface can be connected to RS232 or USB inter face of a PC via electrically isolated interface a...

Page 16: ...dequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 13 2 Disposal instructions Bat...

Reviews: