background image

H40.0.01.6C-09 

Operating Manual GMH 3330 

Page 11 of 16

 

 

Notes to Special Functions 

8.1 

Zero Displacement (’Offset’) 

A zero displacement can be carried out for each of the two temperature channels T1 (TFS0100 only)  
and T2: 

temperature displayed = temperature measured - offset 

Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried out. The zero displacement is mainly 
used to compensate for sensor deviations. Unless 'off' is set, this value will be displayed shortly after the 
device is switched on; during operation it will be identified by means of the Corr-arrow in the display.  

8.2 

Display Correction Factor (’Corr’) 

This factor is applied only to the NiCr-Ni-input T2. 

temperature displayed [°C] = temperature measured [°C] * Corr 

        or temperature displayed [°F] = (temperature measured [°F]-32°F) * Corr + 32°F 

Standard setting: ’off’ =1.000 
This factor is used to compensate for losses of transfer in case of surface measurements, occurring if the 
object to be measured is extremely hot but will be cooled by lower ambient temperatures. The same can 
be true for sensors with a large mass. Unless ’off’ is set, this value will be displayed shortly after the de-
vice is switched on; during operation it will be identified by means of the Corr-arrow in the display. 

8.3 

Base Address (’Adr.’) 

Using the interface converter GRS3105 it is possible to connect several instruments to a single interface. 
As a precondition the base addresses of all devices must not be identical. In case several devices will be 
connected via one interface make sure to configurate the base addresses accordingly.  
Channel 1 will be addressed by the base address set, channels 2 - 6 will have the following addresses. 
 

(Example: base address 21 - channel 1 = 21, channel 2 = 22, 

…, channel 6 = 26) 

How to Calibrate Meas. of Rel. Humidity Using TFS 0100 

Due to the natural aging process of the polymer humidity sensor we recommend to calibrate the sensor at 
least once a year to ensure optimum measuring accuracy. For optimum recalibration and linearity check, 
please return device to manufacturer.  
Use integrated calibration function for 2-point on-site calibration: 

How to calibrate sensor with the calibration device GFN xx

 

 

The following humidity variables are acceptable for the automatic buffer detection: 

Name

 

RH at 20°C

 

Calibration device 

The calibration device GFN XX have been optimized for ap-
plication with TFS 0100. To ensure highly accurate calibra-
tion, we recommend to use these humidity variables only. For 
more detailed information please refer to the relevant operat-
ing manual. 

KNO3

 

93%

 

--- 

NaCl

 

76%

 

GFN 76 

MgCl2

 

33%

 

GFN 33 

Silica-Gel

 

0%

 

---

 

 

Please note: Automatic temperature compensation during calibration  

The rel. humidity to be found in the calibration equipment is quite often highly dependent on temperature. 
This dependence is automatically compensated for when calibrating with the integrated calibration equip-
ment and automatic detection. In case you want to enter calibration values manually, make sure to enter 
the respective temperature with the values.  

 

 

Summary of Contents for GMH 3330

Page 1: ...93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Digital Precision Hand held Measuring Device For Atmospheric Humidity Temperature Dew Point Dew Point Distance Enthalpy and Flow...

Page 2: ...COMBINATION MEASURING SENSOR TFS0100 9 7 MEASUREMENTS USING FLOW MEASURING PROBES STS005 STS020 10 8 NOTES TO SPECIAL FUNCTIONS 11 8 1 ZERO DISPLACEMENT OFFSET 11 8 2 DISPLAY CORRECTION FACTOR CORR 1...

Page 3: ...ombined with the combined measuring probe TFS 0100 is designed for room climate meas urements This includes gauging of atmospheric humidity temperature dew point dew point distance and enthalpy Togeth...

Page 4: ...n lead to malfunctions or destroying of the device and the connected devices This device must not be run with a defective or damaged power supply unit Danger to life due to electrical shock 4 Do not u...

Page 5: ...g 4 1 Display Depending on the measuring probes sensors connected the following measuring results can be displayed TFS 0100 Main display r H relative atmospheric humidity in Secondary display possible...

Page 6: ...ing press Menu for 2 sec configuration menu is activated Store Quit Measurement Hold current measuring value HLD in display for flow meas in AVGHold mode start new measurement or handling of logger fu...

Page 7: ...tc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 Device attached to...

Page 8: ...ll be multiplied by this factor off Factor is deactivated 1 000 Power off Selection of Power Off Delay 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time h...

Page 9: ...ergy con tent of air with a relative humidity of 0 and 0 C is 0kJ kg The warmer the air the higher the relative hu midity the higher the energy content Therefore more energy is required to heat up hum...

Page 10: ...tions tend to be quite high To be able to display a stable measur ing value two averaging functions have been integrated in the instrument Continuous Averaging The average value displayed has been cal...

Page 11: ...sses of all devices must not be identical In case several devices will be connected via one interface make sure to configurate the base addresses accordingly Channel 1 will be addressed by the base ad...

Page 12: ...values a stable value has been de tected and can be taken over by means of the Store key key 6 Then the next calibration step will be displayed 3 Calibration point 2 Put sensor into suitable calibrat...

Page 13: ...cable defective replace Values below measuring range check are there any values below the measuring range specified meas value too low Sensor cable defective replace System fault switch on again if fa...

Page 14: ...Windows 7 Windows 8 8 1 Windows 10 Programming examples Visual Basic 4 0 Delphi 1 0 The following interface functions will be supported Channel Dll Code Name function 1 2 3 4 5 6 x x x x x x 0 Read n...

Page 15: ...lip for table top or sus pended use Weight approx 160 g Interface serial interface 3 5mm jack serial interface can be connected to RS232 or USB inter face of a PC via electrically isolated interface a...

Page 16: ...dequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 13 2 Disposal instructions Bat...

Reviews: