H76.0.02.6C-04
Operating Manual GMH 3251
page 16 of 18
_____________________________________________________
_____________________________________________________________________________
switched on; during operation it will be identified by means of the Corr-arrow in the
display
10.4 Calibration Services
Calibration certificates
– DKD-certificates – other certificates:
If device should be certificated for its accuracy, it is the best solution to return it to the
manufacturer.
Only the manufacturer is capable to do efficient recalibration if necessary to get
results of highest accuracy!
11
Fault and System Messages
Display
Meaning
What to do?
Low battery power, device will only continue
operation for a short period of time
Replace battery
Battery empty
Replace battery
Mains operation without battery: wrong
voltage
Check power supply, replace it when necessary
No sensor connected
Connect sensor
Connected sensor or device defective
If 2nd sensor available, check if device is ok. Return
defective device/sensor to manufacturer for repair
Value extremely out of measuring range
Check: right thermocouple selected.
No display or
confused
characters,
device does not
react on keypress
Battery empty
Replace battery
Mains operation: wrong voltage or polarity
Check power supply, replace it when necessary
System error
Disconnect battery and power supplies, wait shortly,
then reconnect
Device defective
Return to manufacturer for repair
Err.1
Measured value above allowable range
Check: correct thermocouple selected.
Temperature not within sensor range?
-> measuring value to high!
Sensor defective
Return to manufacturer for repair
Err.2
Measured value below allowable range
Check: correct thermocouple selected.
Temperature not within sensor range?
-> measuring value to low!
Sensor defective
Return to manufacturer for repair
Err.3
Display range overflow
Check: value above 1999.9 -> too high to be
displayed
Err.4
Display range underflow
Check: value below -1999.9 (Tara?) -> too low
Er.11
Value could not be calculated
A value, that is necessary for the calculation, is not
available (no sensor) or faulty (over- / underflow
Calculation overflow happened
Choose different unit
Err.7
System error
Check allowed working temperature of the device
Return to manufacturer for repair
12
Reshipment and disposal
12.1 Reshipment
All devices returned to the manufacturer have to be free of any residual
of measuring media and other hazardous substances.
Measuring residuals at housing or sensor may be a risk for persons or
environment