background image

JB02020-INS-LAB-RevA20 • Fingertip Oximeter User Manual 

4

2  IMPORTANT SAFETY GUIDELINES—READ BEFORE USE

The safety statements presented in this chapter refer to 

the basic safety information that the operator of the Finger 

Pulse Oximeter shall pay attention to and abide by. There are 

additional safety statements in other chapters or sections, 

which may be the same as or similar to the following, or specific 

to the operations.

 WARNING: Indicates a potential hazard situation or unsafe 

practice that, if not avoided, could result in death or serious 

injury.

 CAUTION: Indicates a potential hazard or unsafe practice 

that, if not avoided, could result in minor personal injury.

s

  NOTICE: Indicates a potential hazard or unsafe practice 

that, if not avoided, could result in product or property 

damage.

Info: Provides application recommendations or other useful 

information to ensure that you get the most from your product.

Summary of Contents for John Bunn JB02020

Page 1: ...ximeter User Manual Important Do not operate the Finger Pulse Oximeter without first reading and understanding this manual Save this manual for future use Info The most current version of this manual...

Page 2: ...8 CONTENTS PACKING LIST 8 JB02020 FINGER PULSE OXIMETER FEATURES 9 DECLARATION OF CONFORMITY 10 4 SETUP BATTERY INSTALLATION 11 5 OPERATION INSTRUCTIONS 12 CORRECT MEASUREMENT PROCEDURES 13 6 SERVICE...

Page 3: ...tended to measure the pulse rate and functional oxygen saturation SpO2 through a patient s finger It is applicable for spot checking SpO2 and pulse rate of adult and pediatric patients in homes and me...

Page 4: ...e the same as or similar to the following or specific to the operations WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious injury CA...

Page 5: ...o visible damage that may affect patient or user s safety or measurement performance If there is visible damage do not use the device Thereafter inspect the device at least once a week WARNING DO NOT...

Page 6: ...of the oximeter For this reason ensure that all external devices operated in the vicinity of the oximeter comply with the relevant EMC requirements Mobile phones X ray equipment and or MRI devices ar...

Page 7: ...it immediately allow at least fifteen minutes for the Oximeter to reach ambient temperature s NOTICE DO NOT operate the button on the front panel with sharp materials or sharp point s NOTICE DO NOT u...

Page 8: ...er User Manual 8 3 PRODUCT DESCRIPTION AND FEATURES display screen display key rubber cushion JB02020 Finger Pulse Oximeter Contents Packing List 1 One JB02020 Fingertip Oximeter 2 One Lanyard 3 Two A...

Page 9: ...INS LAB RevA20 Fingertip Oximeter User Manual 9 JB02020 Finger Pulse Oximeter Features S L M Display type Displays SpO2 PR PI Auto on off PR and PI shifts Pulse bar Over limits indicator Two direction...

Page 10: ...electrical equipment Part 1 General requirements for basic safety and essential performance BS EN ISO 9919 2009 or the equivalent ISO 80601 2 61 2011 Medical electrical equipment Part 2 61 Particular...

Page 11: ...eries into the battery compartment as shown above Note polarity markings and ensure batteries are correctly installed or device may not function 3 Replace the cover s NOTICE Change the batteries when...

Page 12: ...ace finger between the rubber cushions as shown above and gently close clip on finger 2 After two seconds the Oximeter will automatically power on and begin to measure 3 The display screen is shown ab...

Page 13: ...sh or paint may affect the measuring accuracy and too long a fingernail on measurement finger may cause failure of measurement or inaccurate result Vigorous exercise and electrosurgical device interfe...

Page 14: ...l only or device failure and health hazard could occur Maintenance s NOTICE In order to ensure a long service life adhere to the following maintenance instructions Change batteries when low voltage in...

Page 15: ...e device When the oximeter is taken from a cold environment to a warm and humid environment do not use it immediately allow at least fifteen minutes for the Oximeter to reach ambient temperature Dispo...

Page 16: ...ly as described in OPERATION INSTRUCTIONS try again Unable to turn on device no display 1 Patient must relax as described in OPERATION INSTRUCTIONS try again 2 Change batteries Display direction doesn...

Page 17: ...nm Infrared light 890 nm Maximal average optical output power 2mW SpO2 display range 35 100 SpO2 measuring accuracy 2 for SpO2 range from 70 to 100 Pulse Rate measurement PR display range 30bpm 240bp...

Page 18: ...perating current 40mA Environmental Conditions Temperature Operating 41 F 104 F 5 C 40 C Storage 4 F 140 F 20 C 60 C Humidity Operating 30 80 Storage 10 95 Atmospheric pressure Operating 70kPa 106kPa...

Page 19: ...nally powered equipment Degree of protection against electric shock Type BF applied parts Degree of protection against harmful solid foreign objects and ingress of liquid IP22 with protection against...

Page 20: ...ion Index Pulse Strength Bar Graph Low Battery Voltage CE Mark Serial Number Date of Manufacture Authorized Representative in the European Community Dispose of this Device in Accordance with Local Law...

Page 21: ...the applicable warranty period and which upon examination by GF or its authorized representative prove to be a warranty item This limited warranty is not transferable Within the guidelines set forth i...

Page 22: ...parts installation or other work performed upon or in connection with such products which are not expressly authorized in writing in advance by GF 6 Any labor or shipping charges incurred in the repl...

Page 23: ...ed on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document Some states do not allow the exclusion of certain remedies in those i...

Page 24: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 24 APPENDIX I EMC This equipment meets the requirements of IEC 60601 1 2 2014 Table 1...

Page 25: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 25 Table 1 continued...

Page 26: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 26 Table 2 continued...

Page 27: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 27 Table 2 continued...

Page 28: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 28 Table 2 continued...

Page 29: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 29 Table 3 continued...

Page 30: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 30 Table 3 continued...

Page 31: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 31 Table 3 continued...

Page 32: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 32 Table 3 continued...

Page 33: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 33 Table 3 continued...

Page 34: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 34 Table 4 continued...

Page 35: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 35 Table 4 continued...

Page 36: ...tion can be found on our website www grahamfield com 2020 GF Health Products Inc All Rights Reserved John Bunn is a trademark of GF Health Products Inc One Graham Field Way Atlanta GA 30340 3140 GF He...

Reviews: