Gaussmeter HGM09s Operating Instructions
Chapter 1 Safety Instructions
Page 6 / 75
1.2
1.2
1.2
1.2
Safety Instructions
Safety Instructions
Safety Instructions
Safety Instructions for the Measur
for the Measur
for the Measur
for the Measuring Probes
ing Probes
ing Probes
ing Probes
The magnetic measurements should only be carried out in areas with a max. vol-
tage of 60V DC, 30V AC RMS. The magnetic field probes are not electrically insu-
lated. Please note that the probe holders and the housing might be electrically con-
nected with the protective earth.
If you work in areas with voltages higher than 60V DC, 30V AC RMS or 42V peak
values, act with particular caution because of electric shock hazard.
For measurements in high magnetic fields, please pay attention to the dangers that
may arise by strong magnetic fields.
1.3
1.3
1.3
1.3
Safety
Safety
Safety
Safety Symbols
Symbols
Symbols
Symbols
Safety symbols can be found on various spots on the device.
Before using this connection or function read the corresponding instructions
in the manual.
This symbol refers to information and references in the instruction manual
which the user has to follow in order to avoid injuries to persons or damage
to the device, or to obtain correct measuring results.