Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 59 / 75
7.6.2.6
7.6.2.6
7.6.2.6
7.6.2.6
:RANGe
:RANGe
:RANGe
:RANGe?
?
?
?
Description: Interrogates the current measuring range. 0 = most sensitive range.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :RANG?<LF>
receive 3<CR><LF>
7.6.2.7
7.6.2.7
7.6.2.7
7.6.2.7
:READ?
:READ?
:READ?
:READ?
Description: Displays the current measurement.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :READ?<LF>
receive 2.546313e-01<CR><LF>
7.6.2.8
7.6.2.8
7.6.2.8
7.6.2.8
:UNIT[?]
:UNIT[?]
:UNIT[?]
:UNIT[?]
Description: The physical unit of the measuring value is preset or interrogated.
Mode: Command
Parameter: {
TESL|APM|GAUS|OE|G|T
}
TESL
T
Unit is Tesla
APM
Unit is A/m
GAUS
G
Unit is Gauss
OE
Unit is Oersted
*RST
value: Tesla
Example:
send :UNIT TESL<LF>
send :UNIT?<LF>
receive TESL<CR><LF>