Gossen Metrawatt GmbH
59
•
Measurement Value Storage
•
Clock for Date and Time
•
Functions Monitoring
•
Influencing Quantities and Influence Errors
•
Electrical Safety
Storage Mode
continuous
Memory Capacity,
with 15 min. Interval
with 1 channel:
87,380 records
with 64 channels:
3,971 records
Memory Stability
5 years with backup battery (see also aux.
power supply – backup battery)
Resetting of Meter Values to Zero
via PC or keypad at instrument
Smallest Time Unit
1 s
Allowable Deviation
10 ppm = 5.3 min. per year
Status Display
LED at front panel
Status Relay
changeover contact
Switching Voltage
250 V~, 30 V=
Switching Current
8 A ohmic, 3 A inductive
Switching Cycles
10
5
Influencing Quantity
Nominal Range of Use
Allowable Influence Error
as Percentage of Accuracy Class
Temperature 10
C ... 22 - 24 ... 40
C
0
C ... 22 - 24 ... 55
C
50%
100%
Output Load
load range
20%
HF Interference
IEC 255-4 E5
2.5 kV, 200
,1 MHz, 400 Hz
500%
EM Fields
(severity level 3)
IEC 8001-3
10 V/m 27 - 1000 MHz
500%
EMC Burst
(severity level 3)
IEC 801-4
2 kV, 5/50 ns, 5 kHz
500%
EMC RF cable
(severity level 3)
IEC 801-6
0.15 - 80 MHz, 10 V
200%
Auxiliary Voltage
nominal range of use
10%
Protection Class
I
Overvoltage Category
III
Nominal Insulation Voltage:
Input
Output: Analog, Binary, Uv
Output Relay
Interfaces
Auxiliary Voltage, AC
Auxiliary Voltage, DC
50 V
30 V
250 V
50 V
265 V
80 V
Interference Suppression EN 55022
(VDE 878.3) device: class B
0.15 - 1000 MHz
ESD Protection IEC 801-2
4 kV
EMV Surge (severity level 3)
(IEC 801-5)
2 kV
Summary of Contents for U1601
Page 11: ...Gossen Metrawatt GmbH 11 ECS LAN Sample Network ...
Page 73: ...Gossen Metrawatt GmbH 73 13 Functions Overviews Functions Overview U1601 ...
Page 74: ...74 Gossen Metrawatt GmbH Functions Overview U1602 ...
Page 75: ...Gossen Metrawatt GmbH 75 Functions Overview U1603 ...
Page 79: ...Gossen Metrawatt GmbH 79 ...