12
Gossen Metrawatt GmbH
Reference Ranges
Line voltage
230 V AC
0.2%
Line frequency
50 Hz
2 Hz
waveform
Sine (deviation between effective and rectified value < 0.5%)
Ambient tempe23
C
2 K
Relative humidity
40
60%
Load resistors
Linear
Nominal Ranges of Use
Nominal line voltage
100 V
240 V AC
Nominal line frequency 50 Hz
400 Hz
Line voltage waveform Sinusoidal
Temperature
0
C
+40
C
Ambient Conditions
Storage temperature
-20
C
+60
C
Relative humidity
Max. 75%, no condensation allowed
Elevation
Max. 2000 m
Place of use
Indoors, except within specified ambient
conditions
In order to avoid deviation due to excessive temperature fluctua-
tion, e.g. after outdoor transport at low temperature and subse-
quent operation in a warm indoor environment, it’s advisable to
wait until the test instrument has acclimatized before starting any
measurements.
If the test instrument is colder than the ambient air, condensation
may occur at high humidity, i.e. condensate may accumulate on
components. This could result in the occurrence of parasitic
capacitances and resistances which affect the measuring circuit
and measuring accuracy.
Power Supply
Supply network
TN, TT or IT
Line voltage
90 V
264 V AC
Line frequency
50 Hz
400 Hz
Power consumption
200 mA DUT: Approx. 32 VA
10 A DUT:
Approx. 105 VA
25 A DUT:
Approx. 280 VA
Mains to test socket
(e.g. during function test) Continuous max. 3600 VA, power is con-
ducted through the instrument only,
switching capacity
16 A, ohmic load, the
AT3-IIS32 (Z745X) adapter (for example)
can be used for current > 16 A AC
Electrical Safety
Protection class
I per EN 61140
Nominal voltage
230 V
Test voltage
2.3 kV AC 50 Hz or 3.3 kV DC
(mains circuit / test socket to mains PE ter-
minal, USB, finger contact,
probe(s), test socket)
Measuring category
Designed for 300 V CAT II
(but reduced to 250 V CAT II through the
use of fuses for increased user safety. The
user-friendly fuses are replaceable and
replacements are easily obtainable).
Pollution degree
2
Safety shutdown
At DUT differential current of > 10 mA,
shutdown time: < 500 ms,
can also be set to > 30 mA
in the event of probe current (electronic
fuse) during:
– Leakage current measurement:
> 30 mA~/< 500 ms*
– Protective conductor resistance
measurement:
> 250 mA~/< 1 ms
in case of continuous current I > 16.5 A
Fuse Links
Mains fuses: 2 ea. FF 500V/16A
Special fuse: M 250V/250mA
10 A RPE test current (feature G01 only):
1 ea. FF 500V/16A
Electromagnetic Compatibility
Product standard
EN 61326-1
EN 61326-2-2
USB Data Port
Type
USB slave for PC connection / remote
control**
Type
2 ea. USB master,
for data entry devices * with HID boot
interface,
for USB flash drive for data backup,
for USB flash drive for saving reports as
HTML files,
for printers *
* See section 10 for compatible devices
** Remote control only with extension: “Remote Control via PC (IZYTRONIQ)”
(included as standard feature with SECUTEST ST PRO and available with
SECUTEST DB+ (Z853R or feature KB01).
Bluetooth
®
data interface 2.1 + EDR
(test instruments with feature M01 only)
Frequency range
Max. 2.5 mW (class II)
Transmission intensity 2400 … 2483.5 MHz
Mechanical Design
Display
4.3" multi-display (9.7× 5.5 cm),
backlit, 480 x 272 pixels
at 24-bit color depth (true color)
Dimensions
W × H × D: 295 ×145 × 150 mm
Height with handle: 170 mm
Weight
SECUTEST ST BASE(10)/PRO: approx. 2.5 kg
SECULIFE ST BASE25: approx. 4.0 kg
(depending on the test instrument version)
Protection
Housing: IP 40,
Test socket: IP 20
per EN 60529
Table Excerpt Regarding Significance of IP
Codes
SECULIFE ST BASE(25):
Housing with antimicrobial properties per JIS
standard Z 2801
Database
Number of data
records
50,000
1 data record = 1 DUT or location node or
customer or individual measurement
*
Firmware version 3.2.0 and lower: 12 mA
Interference emission
Class
EN 55011
B
IEC 61000-3-2
B
IEC 61000-3-3
B
Interference
immunity
Test Value *
Evaluation
Criterion
EN 61000-4-2
Contact/atmos. – 4 kV/8 kV
B
EN 61000-4-3
10 V/m (80 MHz … 1 GHz)
A
EN 61000-4-4
Mains connection – 2 kV
B
EN 61000-4-5
Mains connection – 1 kV (LN), 2 kV (LPE)
B
EN 61000-4-6
Mains connection – 3 V
A
EN 61000-4-8
30 A/m
A
EN 61000-4-11
0%: 1 period
B
0%: 250/300 periods
C
40%: 10/12 periods
C
70%: 25/30 periods
C
IP XY
(1
st
digit X)
Protection Against Foreign
Object Ingress
IP XY
(2
nd
digit Y)
Protection Against Water
Ingress
2
12.5 mm
0
Not protected
4
1.0 mm
0
Not protected
Summary of Contents for SECULIFE ST BASE 25