Voltage Measurement
1.
In accordance with the voltage to be measured, turn the rotary switch to V= or V~=
2.
Connect the measurement cables as shown. The connector jack should be grounded.
Make sure that a current measuring range (“A”) has not been activated when the multimeter is connected for voltage
measurement! If the fuse’s blowing limits are exceeded as a result of operator error, both the operator and the
instrument are in danger! With the rotary switch in the V position, the multimeter is always set to the 1 V measuring
range immediately after it is switched on. As soon as the MAN / AUTO key is pressed, and assuming the measured
value is less than 90 mV, the multimeter is switched to the mV measuring range.
Connection of the measurement cables
3