Current Measurement 

 

Notes Regarding Current Measurement: 
 
• The multimeter may only be operated with installed batteries or rechargeable batteries. Dangerous currents are otherwise not indicated, and 
the instrument may be damaged 
• Set up the measuring circuit in a mechanically secure fashion, and secure it against inadvertent breaks. Select conductor cross-sections 
and lay out connections such that they do not overheat 
• An intermittent acoustic signal warns of current greater than 10 A. A continuous acoustic signal warns of current greater than 16 A 
• The input for the current measuring range is equipped with a fuse link. Maximum permissible voltage for the measuring circuit (= rated 
voltage of the fuse) is 1000 V AC/DC. Use specified fuses only! The fuse must have a breaking capacity of at least 30 kA 
• If the fuse for the active current measuring range blows, “FUSE” appears at the digital display, and an acoustic signal is generated at the 
same time 
• If a fuse should blow, eliminate the cause of overload before placing the instrument back into service!  
• Be absolutely certain that the measuring ranges are not overloaded beyond their allowable capacities.  

 

 

Direct and Pulsating Current Measurement, Direct Connection, A DC and A (DC+AC) : 
 

1.

 

First disconnect supply power from the measuring circuit or the power consumer (1), and discharge any capacitors 

2.

 

In accordance with the current to be measured, turn the rotary switch to A= or A~= 

3.

 

Select the current type appropriate for the measured quantity by briefly pressing the FUNC | ENTER multifunction key. 
Each time the key is pressed, the instrument is switched back and forth between A DC and A (DC + AC) TRMS, which is 
indicated by means of an acoustic signal. The current type is indicated at the LCD by means of the DC or the 
(DC+AC)TRMS symbol 

4.

 

Safely connect the measuring instrument (without contact resistance) in series to the power consumer (2) as shown 

5.

 

Switch supply power to the measuring circuit back on (3) 

6.

 

Read the display. Make a note of the measured value if the instrument is not being operated in the memory mode or the 
transmission mode 

7.

 

Disconnect supply power from the measuring circuit or the power consumer (1) once again, and discharge any 
capacitors 

8.

 

Remove the test probes from the measuring point and return the measuring circuit to its normal condition 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for SECULIFE HIT MD

Page 1: ...QUICK SETUP GUIDE SECULIFE HITMD ...

Page 2: ... the finest details Connector for power pack Infrared interface 1 Display LCD MAN AUTO shift key for manual automatic range selection ON OFF LIGHT key for switching the device and the display illumination on and off FUNC ENTER multifunction key Increase measuring range or move decimal point to the right MAN function Rotary switch for measuring functions DAkkS calibration mark DATA MIN MAX key for ...

Page 3: ...orage for more than 15 000 measured values Housing and protective rubber holster with antimicrobial properties Interfaces Bidirectional infrared interface for communication with the PC 38 4 kBd Optionally available IR USB adapter Power supply battery or mains operation via optional broad range variable power pack 95 to 250 V AC IP 52 protection against dust and water Automatic blocking sockets ABS...

Page 4: ...connected for voltage measurement If the fuse s blowing limits are exceeded as a result of operator error both the operator and the instrument are in danger With the rotary switch in the V position the multimeter is always set to the 1 V measuring range immediately after it is switched on As soon as the MAN AUTO key is pressed and assuming the measured value is less than 90 mV the multimeter is sw...

Page 5: ...ltifunction key repeatedly until the unit of measure V or V Fil appears at the display Frequency Measurement 1 Apply the measured quantity in the same way as for voltage measurement 2 Manually select the measuring range for the voltage amplitude When the instrument is switched to frequency measurement the previously selected voltage measuring range remains active 3 You can switch back and forth be...

Page 6: ...ultifunction key repeatedly until unit of measure V or V Fil appears at the display Frequency Measurement 1 Apply the measured quantity is the same way as for voltage measurement 2 Manually select the measuring range for the voltage amplitude When the instrument is switched to frequency measurement the previously selected voltage measuring range remains active 3 You can switch back and forth betwe...

Page 7: ...harge all high voltage capacitors 2 Make sure that the device under test is voltage free Interference voltages distort measurement results 3 Set the rotary switch to the speaker sign as shown as in the picture 4 Connect the conductor path under test as shown Diode Testing with a Constant Current of 1 mA 1 Disconnect supply power from the electrical circuit of the device to be measured and discharg...

Page 8: ...connector cables 000 00 appears at the display After pressing the FUNC ENTER key automatic compensation of cable resistance is activated for all subsequent measurements The short circuit can now be liminated and the device is ready for use Entering Cable Resistance 1 Press the ZERO ESC key once again in the automatic compensation menu 2 Enter the known resistance of the connector cables with the s...

Page 9: ...easured and discharge all high voltage capacitors 2 Make sure that the device under test is voltage free Capacitors must always be discharged before measurement is performed 3 Set the rotary switch to 4 Connect the discharged device under test to the sockets with the measurement cables as shown 8 ...

Page 10: ...solutely certain that the measuring ranges are not overloaded beyond their allowable capacities Direct and Pulsating Current Measurement Direct Connection A DC and A DC AC 1 First disconnect supply power from the measuring circuit or the power consumer 1 and discharge any capacitors 2 In accordance with the current to be measured turn the rotary switch to A or A 3 Select the current type appropria...

Page 11: ...d Hz are alternately selected and switching is acknowledged with an acoustic signal 4 Safely connect the measuring instrument without contact resistance in series to the power consumer as shown 5 Switch supply power to the measuring circuit back on 3 6 Read the display Make a note of the measured value if the instrument is not being operated in the memory mode or the transmission mode 7 Disconnect...

Page 12: ...GMC I Messtechnik GmbH Südwestpark 15 90449 Nürnberg Germany TEL 49 911 8602 111 FAX 49 911 8602 777 www gossenmetrwawatt com info gossenmetrwawatt com ...

Reviews: