26
GMC-I Messtechnik GmbH
10.4
AKTIV - Function Test and Leakage
Current Measurements
Connection
➭
Insert the test object’s mains plug into
the test socket.
➭
Connect the probe to the red probe
socket.
Procedure
➭
Press the
AKTIV
key.
➭
Turn off the device under test.
➭
Switch line voltage to the test socket by
pressing the
OK
key.
As soon as line voltage is applied to the test
socket, the LED for the high-voltage arrow
lights up red.
➭
Switch the device under test on.
➭
Contact any insulated, conductive parts
with the probe.
Polarity Reversal – Devices with Alternating Volt-
age:
➭
Repeat differential and touch current
measurement with reversed polarity:
Press the
key in order to reverse po-
larity (
Reverse L/N
).
Ending the Test
➭
Press the
AKTIV
key in order to end the
test and disconnect the test socket from
the mains.
Standby – Measurement of Small Power Values
Current values of up to 40 mA can be mea-
sured for the display of small power values
(mW range) in standby mode (DUT switched
off).
Procedure
➭
During the function test:
turn off the device under test.
➭
First of all, select the “Standby” function
with the
scroll keys and activate it
with the
scroll key.
The momentary measured value and a mea-
sured maximum value are displayed.
➭
You can repeat the measurement by
pressing the
scroll key (reset). The
max. value is cleared.