10
GMC-I Messtechnik GmbH
9
Data Transmission
9.1
Transmitting Measured Values to the
PC
After pressing the
key, the respective
measured value is initially transmitted via the
Bluetooth interface to an Android device to
which
ELEXONIQ
has been installed (Android
app for
METRALINE PAT
). The data can be
subsequently transferred from the Android
app to a PC, where they can be imported
into
IZYTRONIQ
report generating software.
10
Single Measurements
The following single measurements are
started with the help of the measuring func-
tion keys on the test instrument.
Key
Function
R
SL
/RPE
Protective conductor resistance
measurement
R
ISO
Insulation resistance measure-
ment
I
EA
Equivalent leakage current
measurement
AKTIV
Function test, standby test,
touch current measurement
(direct and differential method),
protective conductor current
measurement
(differential method),
Differential current measurement
(always includes touch current
measurement)
U
Voltage measurement,
probe-probe
Transmitting measured values to
the PC