Calibration Functions: V – Hz
–
Ω
– Temp – mA
– mA
Example of a Periodic Ramp Sequence Controlled by Manual Intervention
Ramp parameters:
output quantity: U (range of 0
…
15 V),
start
= 2 V,
end
= 10 V,
t1
= 5 s,
t2
= 8 s,
t3
= 5 s
,
repeat
for periodic ramp
Key
1 When
ramp
ready
appears at the display:
Start the sequence by pressing
2 Stop the falling ramp within fall time t3 with the
or the
key.
3 Start a rising ramp within remaining fall time t3 with the
key.
4 Stop the ramp sequence by pressing the
or the
key.
5 Start a falling ramp with the
key, remaining dwell time t2 is
cleared.
6 Stop the ramp sequence by pressing
.
6.7.3 Dual Mode (simultaneous simulation and measurement)
All simulation functions can be activated and, at the same time,
transmitter signal U or I can be measured in the most commonly
used measuring ranges (30 V DC or 30 mA DC) in the dual
operating mode.
Special case: simulation and measurement of voltage or current
without an external circuit. For example, if voltage output needs to
be tested, you only need to connect the Cal and Sense+
sockets to each other, because the Calibrator– and Sense– jacks
are connected internally.
Activating the Dual Mode
Ð
Select a simulator function with the rotary switch. The calibra-
tion operating mode must be activated (see section 6.1).
Ð
Press and hold the
ZERO / SEL | ESC
key for approximately 2
seconds, and select the U
dc
or I
dc
measuring function with the
keys.
Ð
Press the
OUT | ENTER
key in order to acknowledge your
selection.
Ð
Depending upon the calibration function, select either the
simulator range or the sensor with the
ZERO / SEL | ESC
key.
Ð
Set the simulation value in the main display with the
keys.
Ð
Activate the output by pressing the
OUT | ENTER
key.
The respective measured values appear in the left-hand auxiliary
display, and
DUAL
appears to the right of the measured values.
Ð
Exit the “dual mode” function by turning the rotary switch, or
by pressing and holding the
ZERO / SEL | ESC
key for
approximately 2 seconds, selecting OFF in the menu with the
keys and acknowledging with
OUT | ENTER
.
2 V
10 V
t1
t2
t3
LCD:
t3
t2
t1
â
ê
ô
û á
é
t1 up
t2 run
t3 dn t3 hld t3 up t3 run t3 hld t1 dn
U [V]
HOLD
CONT
HOLD CONT
/
/
ESC
OUT
ENTER
ZERO /SEL
ESC