GMC-I Messtechnik GmbH
V/Hz,
Ω
, Temperature,
and A/Hz Measurements
Voltage Comparator for Displaying Dangerous Voltage
The input signal or measuring signal is checked by a voltage
comparator for dangerous peaks.
Where U > 33 V AC or U > 70 V DC, two acoustic signals are
generated.
5.2.3 Transient Overvoltages
The multimeters are protected against transient overvoltages of
up to 8 kV with wave-front durations of 1.2 ms and halftimes of
50
μ
s in the voltage measuring range. If longer pulse durations are
expected, for example when conducting measurements at trans-
formers or motors, we recommend the use of our KS30 measur-
ing adapter. It provides protection against transient
overvoltages of up to 6 kV with wave-front durations of 10, and
halftimes of 1000
μ
s. Continuous load capacity amounts to
1200 V
RMS
. Additional influence error caused by the KS30
measuring adapter amounts to approximately -2%.