GMC-I Messtechnik GmbH
49
Calibration Functions: V – Hz
–
Ω
– Temp – mA
– mA
6.5.1 Temperature Simulation for Resistance Thermometers –
RTD Temperature Setting
Resistance temperature detectors (types Pt100, Pt1000, Ni100
and Ni1000) are simulated by means of resistance values.
Abbreviated Instructions
Select calibration function.
Select sensor value and acknowledge for fixed value function.
select
sensor
Pt100
...
ni1000
Set temperature simulator value.
120
.
0
°C
Activate output
:
Response time of the calibrator output to the specified resistance
value after measuring current is applied is max. 30 ms.
Devices under test with non-continuous measuring current (e.g.
scanned measuring inputs) result in erroneous measured values,
if measurement is started before response time has elapsed. The
calibrator cannot be used for devices of this type.
6.5.2 Temperature Simulation for Thermocouples –
TC Temperature Setting
Thermocouples (types B, E, J, K, L, N, R, S, T and U) are
simulated with voltage. Internal or external temperature
compensation is possible.
Abbreviated Instructions
Select calibration function.
Select sensor value and acknowledge for fixed value function.
select
sensor
b
...
u
Set temperature simulator value.
120
.
0
°C
Select internal or external reference temperature,
set external reference temperature (see page 48).
RTD
ON/OFF
LIGHT
ZERO /SEL
ESC
OUT
ENTER
OUT
ENTER
TC
ON/OFF
LIGHT
ZERO /SEL
ESC
OUT
ENTER
Summary of Contents for METRACAL MC
Page 1: ...METRACAL MC Multimeter Calibrator Operating Instructions 3 349 566 03 7 2 15 ...
Page 5: ...GMC I Messtechnik GmbH 5 Contact Persons ...
Page 25: ...GMC I Messtechnik GmbH 25 Control Functions ...
Page 65: ...GMC I Messtechnik GmbH 65 Device and Measuring Parameters ...