GMC-I Messtechnik GmbH
33
GB
10.4 Collection of Used Instruments and Environmentally
Compatible Disposal
The milliohmmeter is a category 9 product (monitoring and
control instrument) in accordance with ElektroG
(German
Electrical and Electronic Device Law)
. This device is sub-
ject to the RoHS Directive. Further-more, we make refer-
ence to the fact that the current status in this regard can
be accessed on the Internet at www.gossenmetra-
watt.com by entering the search term WEEE.
We identify our electrical and electronic devices in
accordance with WEEE 2012/19/EU and ElektroG
with the symbol shown to the right per DIN EN
50419. These devices may not be disposed with
the trash. Please contact our service department regarding
the return of old devices. (see chapter 12).
If you use type-tested
batteries
in your instrument or acces-
sories which no longer function properly, they must be duly
disposed of in compliance with the applicable national reg-
ulations.
Batteries or rechargeable batteries may contain harmful
substances or heavy metal such as lead (PB), cadmium
(CD) or mercury (Hg).
They symbol shown to the right indicates that
batteries or rechargeable batteries may not be
disposed of with the trash, but must be delivered
to collection points specially provided for this
purpose.
11
System Messages
The following messages appear at the main display, or the
auxiliary displays as required. See “Symbols used in the
Digital Display” on page 3 for messages displayed at visible
segments.
Blinking Unit of Measure
All measuring functions are balanced for each METRA
HIT
⏐
27EX
at the factory in accordance with the respective
technical specification. If a unit of measure blinks, this indi-
cates that balancing constants which have been estab-
lished and saved to the milliohmmeter are no longer avail-
able for the respective function. If this is the case, mea-
surement results may deviate from the specification. We
recommend sending the instrument to our Repair and
Replacement Parts department for rebalancing (see chap-
ter 12).
Message
Function
Meaning
0
.
L
Measuring
Indicates overflow
open
L
4-wire-m
Ω
Measuring current interrupted or fuse defective.
Pb Cd Hg