7.2
Thermovoltage Compensation
Thermovoltages which occur as a result of material and
temperature differences may distort measurement results.
For this reason, the instrument is equipped with automatic
thermovoltage compensation in the relevant measuring
ranges.
After switching on the milliohmmeter, „
termoc
“ is displayed,
indicating that an automatic thermovoltage compensation
for the 30 m
Ω
and 300 m
Ω
measuring ranges will be per-
formed in the background for all future measurements.
ZERO signals that the value of the thermovoltage has
already been taken into account for the indicated measu-
ring result.
After the first measurement – lift the test probe from the
measuring point – „
termoc
“ disappears to facilitate the rea-
ding of the measured values. The automatic thermovoltage
compansation remains activated.
000
.
10
termoC
m
m
Ω
Ω
Measure
thermo-
voltage
Relevant measuring ranges:
30 m
Ω
und 300 m
Ω
I < 100 mA
ZERO