10
GMC-I Messtechnik GmbH
4
Selecting Measuring Functions and Measuring
Ranges
4.1
Automatic Measuring Range Selection
The milliohmmeter is equipped with auto-ranging for all
measuring ranges. Auto-ranging is active as soon as the
instrument is switched on. The instrument automatically
selects the measuring range which allows for highest pos-
sible resolution for the applied quantity.
The instrument automatically switches to the next highest
or next lowest measuring range for the following measured
quantities:
4.2
Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges
can be selected manually in accordance with the following
table.
The manual mode is deactivated by pressing and holding
the
MAN|AUTO
key (approx. 1 s) or by switching the instru-
ment off and then back on again.
Note!
If the red ERROR LED lights up and OL is shown
on the display during manual measuring range
selection, you should switch to the next higher
range.
4.3
Quick Measurements
If you wish to perform quicker measurements than those
possible with the automatic measuring range selection
function, make sure to establish the appropriate measuring
range and to perform the following two functions:
•
by
manual measuring range selection
, i. e. by selecting
the measuring range with the best resolution, see
chapter 4.2.
or
Measuring Ranges
Reso-
lution
Switching to next
highest range
at
+
(... d + 1 d)
Switching to next
lowest range
at
+
(... d –1 d)
300 m
Ω
, 3
Ω
, 30
Ω
4 ¾
31 000
2 800
30 m
Ω
3 ¾
3 100
280
MAN|
AUTO
Function
Acknow-
ledgement
Dis-
play
Acoust.
Signal
short
Manual mode active:
utilized measuring range is fixed
MAN
1 x
short
Range switching sequence for:
m
Ω
: 30 m
Ω →
300 m
Ω →
3
Ω →
30
Ω →
30 m
Ω → …
MAN
1 x
long
Return to automatic range selection
—
2 x