SUGGESTED PULSE SETTINGS

The  chart  below  lists 

approximate

  pulses  necessary  to  penetrate  various  materials.    Settings  vary  greatly 

depending on imaging system used.  Refer to imaging system instructions for more information.  

 

  

MATERIAL

PULSE SETTING 

(24” BETWEEN X-RAY & IMAGER)

CARDBOARD / LIGHT WOOD / PLASTIC

2-5

LIGHT METAL

10

STEEL 1/4”

25

STEEL ½” 

50

STEEL 1”

99

X-ray Beam

25’ (7.6 m)

25’ (7.6 m)

100’ (30m)

10’ (3m)

10’ (3m)

10’ (3m)

Figure 4: Exclusion Zone

OPERATING PRECAUTIONS:  

The operator should always stand at least 10 feet (3m) behind the X-ray unit and clear 

all personnel at least 10 feet (3m) behind the unit or at least 100 feet (30m) from the front of the unit before pulsing.  
The exclusion zone (below) should be a controlled area free of all personnel while X-ray pulses.   

*XRS3

 

should be positioned directly in front of the object to be X-rayed and the imager placed directly behind

the object to be X-rayed.  Imager should be placed as close to the object as possible.  Distance between XRS3 and
imager is usually 24 to 72 inches (30 to 180 cm).  During operation XR200 should be stabilized on a   at surface,
a tripod, or a custom   xture suitable for holding the 11 pound (5 Kg) XRS3.

24 to 72 inches (30 to 180

 cm)

OBJECT

IMAGER

XRS3

Summary of Contents for XRS-3

Page 1: ...OPERATOR S MANUAL MARCH 2018 Golden Engineering Portable X ray Technology...

Page 2: ...ERATING PRECAUTIONS 8 EXCLUSION ZONE 8 PULSE SETTINGS 8 6 0 NAVIGATING THE MENU 9 7 0 MAINTENANCE 11 DOSE MEASUREMENT 11 TUBE REPLACEMENT 11 8 0 TROUBLE SHOOTING 11 9 0 INSTRUCTIONS FOR SERVICE 12 REM...

Page 3: ...dge and or a pocket dosimeter consistent with the appropriate federal territorial or provincial standards note an electronic dosimeter will not detect the XRS3 radiation pulses Due to the short pulse...

Page 4: ...mount The identi cation label located on the bottom of the generator lists the manufacturer s name and address model number serial number weight volt amp and production date BATTERY PACK The standard...

Page 5: ...igit LCD Home displays number of pulses selected and delay time before the rst pulse Home will also indicate if multiple pulse trains have been entered Arrows allow operator to scroll through various...

Page 6: ...ABLE CONNECTOR Lemo K series ve pin connector located on the back of the control module beneath the battery receives the remote cable or imaging system cable POSITIVE REMOTE CABLE CONNECTOR BATTERY TE...

Page 7: ...ulsed 7 As the High Voltage Switch is closed a high voltage transient of approximately 270 000 volts and 25 nanoseconds in duration is applied across the x ray tube generating x rays The closing of th...

Page 8: ...7 5 0 OPERATING INSTRUCTIONS The following are basic operation instructions to take an X ray image using the XRS3 Certain applications may require modi cations to these basic procedures...

Page 9: ...feet 3m behind the X ray unit and clear all personnel at least 10 feet 3m behind the unit or at least 100 feet 30m from the front of the unit before pulsing The exclusion zone below should be a contro...

Page 10: ...pulse setting LCD ERROR MESSAGE Low Battery Battery voltage is too low to pulse the XRS3 Charge the battery PWM not running Could occur at the start of a pulse train or during a pulse train Power uni...

Page 11: ...2 _ _ _ _ RESET TRAINS SETTINGS BACK HOME MENU PRESETS SETTINGS BACKLIGHT SET PULSE LIFE PC BACKLIGHT FEEDBACK ON OFF SET DELAY RESET PC X RAY INFO TRAINS SETTINGS BACK SETTINGS FEEDBACK BACKLIGHT FEE...

Page 12: ...the tube s output will decrease slowly with use If the tube is broken or the glass cracks the tube output will cease immediately 8 0 TROUBLESHOOTING SYMPTOM TEST ACTION No power on light Check batter...

Page 13: ...he control module 3 Remove cable connecting the boards 4 Control Module Removed 1 2 3 4 Remove keyswitch connector Remove 3 screws holding processor board to top Remove processor board Disconnect ribb...

Page 14: ...ited warranty customer must ship the entire unit or the component believed to be defective to Golden Engineering post paid Golden Engineering assumes no liability for units or components shipped until...

Page 15: ...onds ELECTRICAL AND THERMAL CHARACTERISTICS Battery voltage 18 volts 20 volts Battery type Li Ion Battery recharge time One Hour Current draw 20 amps 18 20 volt battery Current ow 0 25 mA Temperature...

Page 16: ......

Reviews: