B l u e C o n n e c t – M o d b u s - A d d r e s s e s S e n s o r M o d u l e s
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 23 / 26
www.go-sys.de
BlueConnect Redox 486 C000-9 – Modbus Address Overview
31.8.2021
Address Parameter name
Range
Meaning
Data type
Authorisation
0x00
Device ID
106
106 – BlueConnect Redox
Short
R
0x01
Firmware Version
100 – 9999
100 = 1.00, 2410 = 24.1
Short
R
0x02
Serial No.
0 – 65535
Serial number
Short
R
0x03
Modbus Slave ID
1 – 230
Modbus Address
Short
R/W
0x04
Baud rate
0 – 2
0 = 9600 8N1
Short
R
0x05
Production date
ddmmyyyy
Date
Short x 2
R
Address Parameter name
Range
Meaning
Data type
Authorisation
0x14
A0
0 – 0xffffffff
Cal Coefficient A0
32 Bit Float
R/W
0x16
A1
0 – 0xffffffff
Cal Coefficient A1
32 Bit Float
R/W
0x18
A2
0 – 0xffffffff
Cal Coefficient A2
32 Bit Float
R/W
0x1A
A3
0 – 0xffffffff
Cal Coefficient A3
32 Bit Float
R/W
0x1C
A4
0 – 0xffffffff
Cal Coefficient A4
32 Bit Float
R/W
0x1E
A5
0 – 0xffffffff
Cal Coefficient A5
32 Bit Float
R/W
Address
Parameter name
Range
Data type
Authorisation
0x101
Redox [mV]
0 – 0xffffffff
32 Bit Float
R
0x104
Temperature [°C]
0 – 0xffffffff
32 Bit Float
R
Note on 32 bit Float data
(MSB = 0xByte4, LSB = 0xByte1),
The Receiving sequence of the values (Hex) is: 0x [Byte2] [Byte1] [Byte4] [Byte3]