I S A – A M S O p e r a t i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 38 / 76
www.go-sys.de
9.3 The Sensor Setup Window of an Application-Specific Parameter
The application-specific parameters generated from spectral data are, such as virtual sensors (see
13 Virtual
Sensors
), calculated parameters. The main difference to virtual sensors is the type of calibration that allows a
continuous calculation of the
SQI
1
(Spectral Quality Index).
AMS handles the application-specific parameters such as sensors.
By default, each spectrometer is configured with SAC254 as an application-specific parameter. In addition,
there are three free parameters which can be customized configured. For more information on customer
specific application parameters, please contact GO Systemelektronik.
Example COD:
COD
Double-click in the AMS start window
This opens the sensor setup window. The factory default values for the spectrometer are factory preset.
The CAN-ID of a newer spectrometer starts with
ISA
in capital letters, e.g.
ISA00001
.
Opens the configuration window of the parameter.
see
9.
3.1 The Configuration Window of an Application-Specific Parameter
Opens a window with the display of the last recorded single value of an
averaged measurement value.
3
1
SQI (Spectral Quality Index) see Page 41 top and
Appendix D – SQI (Spectral Quality Index)
2
CAN-ID + sensor number (uniquely defined for each sensor, factory preset)
3
Is useful in some cases for testing purposes.
Formula input field
The field is inoperable here.
Last measured parameter value
with measurement time
Sensor-ID
2
of the parameter
Display
SQI value