CORRECTIVE AND PREVENTIVE MAINTENANCE AND
PRESERVATION
• WATER FILTER*
Disarm the filter as the picture
below. Wash the sintered with brush
and dry with compressed air.
When mounting again, verify
the filter direction, water must flow
through the filter on the arrow
direction of the following picture.
- Use clean and soft cloth dampened in alcohol 70% to disinfection of the equipment.
- Never use corrosive disinfectants or solvents.
Disinfection
Note: Use gloves and other
systems of protection,
during the disinfection.
Sterilization
*Transducer sterilization:
Remove the insertion from the transducer.
Carefully remove the transducer (01) from the adapter
(02) by means of pressure, “do not try to rotate
the piece”; afterwards take it for sterilization in the
autoclave (packed).
01
02
Autoclavables:
*Removable transducer, insertions
and key are autoclavables on the
following conditions:
- Maximum temperature of 135ºC
46
*
Optional item
The equipment must suffer routinely measurements, following the current legislation
of the country.
But, never with a period superior to 3 years.
For protecting your equipment, look for a Gnatus’ technical assistance for periodic reviews
as preventive maintenances.
Preventive Maintenance
Gnatus states that the supplying of the circuits’ diagram, Part lists or any other information
that permits the technical assistance by the user, can be requested, since previously agreed
between the buyer and Gnatus.
Corrective Maintenance