In order to avoid environmental contamination or undue use of the Equipment after it
has become useless, it should be discarded in the suitable place (as per the local legislation
of the country).
- Pay attention to the local legislation of the country for the conditions of installation
and disposal of residue.
Precautions to be adopted against foreseeable or uncommon
risks, related to the deactivation and abandoning of equipment
- If the equipment has any abnormality, check if the problem is related to any item
listed in the topic of unforeseen events (failures, causes and solutions). If it is not
possible to resolve the problem, turn off the equipment, remove the power supply cable
from the socket and contact your representative (Gnatus).
Precautions in case of alteration in the functioning of equipment
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing
to be cleaned and disinfected.
CORRECTIVE AND PREVENTIVE MAINTENANCE AND
PRESERVATION
• BODY OF THE EQUIPMENT, CABLE, TRANSDUCTOR AND HOSES
To clean the equipment, we recommend the use of
“BactSpray”
(Reg nº
MS: 3.2079.0041.001-5) or any other similar product: Active component:
Benzalkonium chloride (tri-quaternary ammonium)
Solution 50%................................................. 0.329%
Chemical composition:
Butyl Glycol, Decyl polyglucose, Sodium Benzoate,
Sodium Nitrate, Essence, Deodorized Propane / Butane, demineralized Water.
For more information concerning cleaning procedures, see manufacturer’s
instructions.
NOTE:
The registration at the Ministry of Health of the “BactSpray”
is executed separately from the product described in this manual, as the
“BactSpray” is not manufactured by Gnatus.
Cleaning
Important:
In order to execute cleaning or any type of maintenance, ensure that the
equipment is disconnected from the electrical network.
PRECAUTIONS, RESTRICTIONS AND WARNINGS
45
WARNING:
- In order to prevent risks and damages to equipment, make sure that the liquid does
not enter into the unit.
- The application of other solvent-based cleaning products or sodium hypochloride isn’t
recommended, because they may damage the equipment.
The cleaning procedure should be executed at the start of the
working day and after each patient. Always turn the main switch
off before executing the procedures of daily maintenance.
Summary of Contents for EasySonic
Page 1: ......
Page 7: ...MÓDULOS ACESSÓRIOS OPCIONAIS E MATERIAIS DE CONSUMO 07 01 02 03 04 05 06 07 08 09 10 11 12 ...
Page 54: ...MÓDULOS ACCESORIOS OPCIONALESY MATERIALES DE CONSUMO 54 01 02 03 04 05 06 07 08 09 10 11 12 ...
Page 73: ...73 ...
Page 74: ...74 ...
Page 75: ...75 ...
Page 76: ......