TECHNICAL SPECIFICATIONS

Product classification:

According to norm NBR IEC
60601-1

Supply voltage

127/220 V~ (manual selection)

Frequency

50/60 Hz

Electric shock protection
type:

Class one equipment.

Degree of protection
against electric shock:

B type

Operation way

Continuous with intermittent
load: T-on 1min. - T-off 4min.

Water leak protection

IPX0 – All the chair,
excepting foot control
IPX1 – foot control

• Power

See table above

Protection fuses

F1 e F2 ...........8A - Time-lag

Rising capacity

200 Kg

• Delivery unit tray’s maximum

load capacity

Syncrus line H - 1 Kg
Sincrus lines GL and L - 2 Kg

Net Weight

115 Kg

Gross Weight

154 Kg

FREQUENCY

CONSUMPTION

NOMINAL TENSION

POWER

127V~

220V~

220V~

60Hz

60Hz

50Hz

4,2A

2,4A

2,4A

530VA

530VA

530VA

Special providence or particular conditions for installation

- Check the electric net, it must be compatible with the specified one in the equipment.
- Verify the thread earth, it must be turned off correctly.
- Verify the general key of the equipment in the position “0”.

OBS:

 These information also make part of the Manual of Installation and Maintenance of the equipment

that can be found with the authorized Gnatus technician.

Transport and storage conditions

The equipment must be transported and stored with the following observations:
- Carefully, should not suffer drop and neither receive impact.
- With the side of the arrow upward.
- With humidity protection, not to expose to rains, sparkling of water or humidified floor.
- With temperatures from -12ºC to 50ºC.
- With maximum piling up of 04 units for storage (DENTAL CHAIR).
- With maximum piling up of 07 units for storage (UPHOLSTERY).

Standards

This equipment was designed, manufactured and tested in compliance with the following standards:
NBR-IEC série 601-1 Equipamento Eletromédico - Parte 1: Prescrições gerais para segurança;
NBR ISO 6875:1998 - Equipamento odontológico – Cadeira odontológica de paciente;
EN 980:2003  (Ed. 2)  - Graphical symbols for use in the labelling of medical devices;
ISO 14971 - Medical devices - application of risk management medical devices;
ISO 9687: 1993 -  Dental equipment - graphical symbols;
ISO 13485-2 - Quality systems - medical devices;
ISO 780 - Packaging - pictorial marking for handling goods.

• Electromagnetic compatibility

This product was tested and approved in compliance with the following standards:

EN 60601-1 (1990);
Amendment 1 EN 60601-1 (1992);
Amendment 2 EN 60601-1 (1995);
Amendment13 EN 60601-1 (1995);
EN 60601-1-2 (2001);

NBR IEC 601-1 (1994);
Emenda 1 NBR IEC 601-1

(1994);

NBR IEC 601-1-2 (1997);
CISPR 11, edição 3.1 (1999);

IEC 61000-4-2 (1999);
IEC 61000-4-3 (1998);
IEC 61000-4-4 (1995);
IEC 61000-4-5 (1995);
IEC 61000-4-6 (1996);
IEC 61000-4-11 (1996);

Summary of Contents for Cadeira Syncrus GLX

Page 1: ......

Page 2: ...nty of the Equipment 15 General Data 15 Technical Specifications 17 Symbols 18 Dimensions 19 Operation 20 Maintenance 23 Precautions 23 Cleaning 23 Troubleshooting 24 Final Considerations 24 Descripción del Producto 25 Garantía del Equipamiento 26 Datos Generales 26 Especificaciones Técnicas 28 Simbolos 29 Dimensiones 30 Funcionamiento 31 Mantenimiento 34 Precauciones 34 Limpieza 34 Imprevistos 35...

Page 3: ...t productivity avoiding unnecessary movements and making cleaning and disinfections much easier Optional second armrest with side opening designed to ease the patient s access Ample seamless upholstery with lumbar rest mounted on a rigid structure covered with a high resistant polyurethane cover this coverage use material plastic and seamless upholstery Removable Backrest easy operating the Easy F...

Page 4: ...ice Always having in mind the dentist s security we developed the leg bump preventing device When the chair were descending any obstruction below the chair would trigger this device interrupting all functions and preventing further complications see page 20 Electronic board with reversion of voltage The board of electronic circuit allows tHhe connection in any tension of entry 127V or 220V 50 60Hz...

Page 5: ... Swivel armrest with side opening optional 14 Headrest knob 15 Cervical rest optional 01 02 03 04 05 06 07 08 09 10 11 12 14 13 15 List of pieces and circuit scheme Gnatus Company declares that the supply of the circuit scheme list of pieces or any other information that propitiate technical attendance for the user can be request if there is an agreement between the user and Gnatus Company ...

Page 6: ...st be transported and stored with the following observations Carefully should not suffer drop and neither receive impact With the side of the arrow upward With humidity protection not to expose to rains sparkling of water or humidified floor With temperatures from 12ºC to 50ºC With maximum piling up of 04 units for storage DENTAL CHAIR With maximum piling up of 07 units for storage UPHOLSTERY Stan...

Page 7: ...f the dental light high density Activation of the dental light low density It determines the initial position It determines to spitting position last position Lift backrest Lower backrest Lift seat Lower seat It determines the work position 1 If determines the work position 2 B type equipment Turned on position Emergency stop Turned off position Landing in many parts of the equipment indicates the...

Page 8: ...English 19 Dimensions mm TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations ...

Page 9: ...per the procedure below Poisitioned in 220V Poisitioned in 127V 1 Desconnect the general key of the chair 2 Remove the finishing cover of the motor 3 Localize the Electronic Board removing the protective cover 4 Put the Jumper 33 accoring to the voltage of the local electric net 127V 220V 5 After doing the operation assembly the covers and turn on again the equipment to the electrical net 33 ATTEN...

Page 10: ...r minutes and the backrest will go up and the set will go down To switch the dental light on press the key 20 or 22 according to desired intensity to change intensity just press the other key To switch it off press the key again After pressing the Inital position 19 key or the Last position Spitting position 18 key any other operation will trigger the Stop and automatically the backrest current po...

Page 11: ...tion below the chair would trigger this function interrupting all movements and preventing further complications Swivel seat The Syncrus GLX chair can swivel thus making cleaning and disinfection easy To move it hold the seat end as shown in picture Headrest To move the headrest loose knob 14 turn it anticlockwise until desired position is reached and fix it turning it clockwise To adjust height j...

Page 12: ...and humidity When moving the chair never get it for the part of the support of the legs lift it for the structure in the area of the seat The plastic parts can t be in contact with most chemical substances used in dentistry treatments i e acids mercury acrylic liquids amalgams etc Whenever off duty turn off the main switch Replacing the fuse With the aid of a screw driver loosen the fuse holder co...

Page 13: ...r the correct maintenance of your equipment Doubts and information GNATUS Call center 55 16 2102 5000 ATTENTION Any other problem that could happen with its equipment should only be repaired by an authorized technician Gnatus The bad use negligence or maintenance not made by technician authorized by the maker will imply in the loss of the warranty Imprevist Problem Probable Cause Solution Chair is...

Page 14: ......

Reviews: