The use of PC and Printer in the Patient Environment is defined below and must be followed
to ensure safety compliance of the instrument
Option 1: Patient Environment for Non
–
Medically approved PC & Printer
Patient Environment (3 metres)
Figure 1: Patient Environment for Non-Medically approved PC & Printer
If the PC and/or printer are non-medically approved then it should be positioned as shown in
Figure 1, out with the patient environment. The PC and/or printer must be compliant to
EN60950-1.
Option 2: Patient Environment for Medically approved PC & Printer
Figure 2: Patient Environment for Medically approved PC & Printer
If the PC and/or printer is compliant to BS EN 60601-1 and medically approved then the
Printer and PC can be positioned within the patient environment as shown in Figure 2.
In circumstances where the PC and /or printer are not medically approved but are to be used
within the patient environment then an isolation transformer, which is in compliance to BS
EN 60601-1 MUST be used to ensure safety compliance, or run the PC by only battery
without mains connection.
NR6
Rhinomanometer
Flowhead Tubing
(1 metre)
USB cable
(2 metre)
NR6
Rhinomanometer
USB cable
(2 metre)
Flowhead Tubing
(1 metre)
Patient Environment (3 metres)