Rhinomanometer NR6
NR6 User
’s
Manual V 16 14/02/2020 P28 of 34
Appendix 1: Electromagnetic Compatibility
Electromagnetic Compatibility Requirements - Emissions
Required Test
Compliance
Comments
RF emissions
CISPR 11
Group 1
Class B
For use in a professional
healthcare facility environment
only
Harmonic emissions
IEC61000-3-2
N/A
Voltage fluctuations / flicker
emissions
IEC61000-3-3
N/A
Electromagnetic Compatibility Requirements - Immunity
Required Test
60601 test level for equipment
used in a professional
healthcare facility only
Compliance level
Electrostatic discharge (ESD)
IEC61000-4-2
± 8 kV contact
± 2, 4, 8, 15 kV air
± 8 kV contact
± 2, 4, 8, 15 kV air
Radiated RF EM Fields
IEC61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
80 MHz to 2.7 GHz
Proximity fields from RF wireless
communications
IEC61000-4-3
385MHz - 27 V/m
450MHz - 28 V/m
710, 745, 780MHz - 9 V/m
810, 870, 930MHz - 28V/m
1.72, 1.845, 1.97GHz - 28 V/m
2.45 GHz - 28 V/m
5.24, 5.5, 5.875GHz - 9 V/m
385MHz - 27 V/m
450MHz - 28 V/m
710, 745, 780MHz - 9 V/m
810, 870, 930MHz - 28V/m
1.72, 1.845, 1.97GHz - 28 V/m
2.45 GHz - 28 V/m
5.24, 5.5, 5.875GHz - 9 V/m
Electrical fast transient / burst
IEC61000-4-4
± 2 kV for power supply lines
± 1 kV for input / output lines
N/A
Surge
IEC61000-4-5
± 0.5, 1, 2 kV line(s) to earth,
± 0.5, 1 kV line(s) to line(s) for
power supply lines
± 2 kV line(s) to earth for input /
output lines
N/A