A1 ACOUSTIC RHINOMETER
CALIBRATION
The Acoustic Rhinometer should be calibrated each day using the built-in facility. In addition
to this, the cables should be inspected regularly for signs of damage, in particular the region
of the sound tube and its link to the PC.
The nosepieces supplied are for single-use only. Care should be taken in use to prevent the
passage of nasal secretions down the nosepiece and into the sound tube. Should this occur,
the sound tube should be removed from the sound box, cleaned by one of the methods
described below and then thoroughly dried prior to reconnecting it to the sound box.
N.B.
Take care not to over tighten the fixing screws when re-fitting the tube as this could
damage the tube.
ROUTINE MAINTENANCE
Nosepieces are supplied for single use only.
Care should be taken in use to prevent the passage of nasal secretions down the nosepieces
and into the sound tube. Should this occur, the sound tube should be removed from the
sound box, cleaned by one of the methods described below and then thoroughly dried prior
to fitting back onto the sound box.
N.B.
take care to not over tighten the fixing plastic screws when re-fitting the tube as this
may cause damage to it