Naris
Naris
User’s
Manual V14 14/02/20 Page 31 of 52
Scan channels
= CH2 through CH2
A/D range
= ± 10 Volts
Scan rate
= 1000
Mode
= Continuous
Data format
= Volts
Click on OK/Start.
If you now create a pressure on the NR6 pressure tube, or gently press on the pressure
nozzle the trace will show this. The trace should lie at close to zero units and travel up and
down from there. If it lies significantly lower or higher than zero units the
Pressure Offset
potentiometer on the NR6 circuit board (VR4) can be adjusted in real time to correct this.