CLEANING OF FLOWHEAD
Due to the nature of the flowheads then Autoclaving, High Energy Irradiation
and Boiling Water are NOT allowable
.
Should cleaning become necessary then we suggest the following cleaning/disinfection
process.
Manual Cleaning/ Disinfection
1.
Prepare a 2% (30ml/l) cleaning and disinfection of Sekusept® AKTIV with deionised
water at 20 °C (68°C)
2.
After 15 minutes the cleaning and disinfection solution can be used
3.
Clean the flowhead with soft sponges in the cleaning and disinfection solution. Any
areas difficult to access should be reached with soft brushes.
4.
Leave the flowhead in the solution for 15 minutes ensuring that it is fully submerged
5.
Remove from the cleaning/disinfection solution and rinse thoroughly with deionised
water
6.
Dry the device thoroughly
7.
Check for visible contamination and repeat steps above if required
8.
Check the flowhead for damage
Special care should be taken to ensure that the annular rings, which connect to the pressure
ports within the flowhead, are also thoroughly dry before re-using the head. Washing does
not affect the pressure drop produced at a given flow and therefore does not alter the
instrument calibration, provided all detergent etc is removed before the heads are
thoroughly dried
The flowheads can be repeatedly sterilised without limitation as long as there is no physical
damage observed to any of the parts. If damage is observed and/or measurements become
unusual then please contact to GMI to arrange a replacement.
Note: Due to the nature of the flowhead then it is possible to use Ethylene Oxide Processing,
if you have this facility available to you and you wish to use it. We recommend that you
follow the protocol that is established at your facility and should there be any issues
identified then please contact to GMI for a replacement.
Note definitions of Definitions for Cleaning/Disinfecting/Sterilising are show in Appendix 4