NARIS SAFETY INFORMATION
NR6 RHINOMANOMETER SAFETY INFORMATION
Read this Operating Manual before attempting to use the
Instrument.
WARNINGS
This instrument is for indoor use only and it should be used as described in this manual.
The system must not be used in the presence of flammable gases or in an environment which
is susceptible to explosions (Beware of oxygen, dust, and anaesthetic gases)
To avoid risk of electric shock, the PC, if connected to a mains supply must be connected to a
mains supply with a protective earth.
The unit is a PC-connected product. It is advisable not to touch the patient whilst using the
equipment.
The equipment should be positioned in such a way that it can be easily disconnected from a
mains-powered PC. The operation of the system can be safely terminated by switching off the
PC, or removing the USB cable between the PC and the NR6
If your PC or printer does not have a power supply approved for a patient environment, then
an isolation transformer, which complies with BS EN 60601-1, should be used to power the PC,
Printer. You must use the transformer or run the PC by battery to ensure that the NR6 is in
compliance with BS EN 60601-1.
Applied Parts. The applied parts consist of single-use foam inserts, microfoam tape, tip
connectors, anterior or posterior tubing, masks (single-use /or reusable) connected to an
antiviral filter, flowheads and silicon rubber tubing.
CAUTIONS
Patient connection components may cause an irritation reaction in some patients. Use of
such components should be discontinued in patients who exhibit such a reaction.
Certain components are identified as single-use items. Single-use items should not be
reused as they could carry infections between subjects
Federal (USA) law restricts this device to sale by or on the order of a physician.
The use of an NR6 Rhinomanometer near to sources of electromagnetic radiation, such as
mobile phones, radio transmitters, x-ray equipment etc, may prevent it from functioning