1. To Set the Output Load
For the [Output] connector on the front panel, the generator has a built-in
50
Ω
series impendence. If the actual load does not match the set one, the
displayed amplitude and offset will be incorrect. This function is used to
match the displayed voltage with the expected one.
Steps for setting the load:
Press Utility
→
Output Setup
→
Load, to enter the following interface.
Please note that the load parameter shown on the right bottom
is the default setting when the power is on or the pre-set load value. If the
current value is valid for the output, then current value will be used.
Figure 2- 59 Set the Output Load
Instruction
SFG-20X Series has a fixed 50
Ω
Series Impendence. No matter what value the
set parameter is, if the real load is different from the set one, the displayed voltage