6
globalspecialties.com
Global Specialties
Do not Operate
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other
material gets on or inside the instrument, remove the instrument’s power cord, remove the
instrument from service, label it as not to be operated, and return the instrument to Global
Specialties for repair. Notify Global Specialties of the nature of any contamination of the
instrument.
Not for Critical Applications
This instrument is not authorized for use in contact with the human body or for use as a
component in a life-support device or system.
Instrument covers must not be removed by operating personnel. Component
replacement and internal adjustments must be made by qualified service
trained maintenance personnel who are aware of the hazards involved when
the instrument’s covers and shields are removed. Under certain conditions,
even with the power cord removed, dangerous voltages may exist when
the covers are removed. To avoid injuries, always disconnect the power
cord from the instrument, remove all other connections (for example, test
leads, probes, etc.), discharge all circuits, and verify there are no hazardous
voltages present on any conductors by measurements with a properly
operating voltage-sensing device before touching any internal parts. Verify
the voltage-sensing device is working properly before and after making the
measurements by testing with known operating voltage sources and test
voltages. Do not attempt any service or adjustment unless another person
capable of rendering first aid and resuscitation is present. Hazardous
voltages may be present in unexpected locations in circuitry being tested
when a fault condition in the circuit exists.
Use Correctly Sized Wires
Do not Touch Live Circuit
Use only standard 4 mm banana plugs when connecting to power supplies.
Solid core wires no larger than 20 AWG must be used for connections to the
solderless breadboards .
WARNING
CAUTION
Do not Attempt Servicing