Electrical Power
This instrument is intended to be powered from a CATEGORY II mains
power environment. The mains power should be 115 VAC or 230 VAC
(±10%) at 50/60 Hz. Use only the power cord supplied with the instrument
and ensure it is appropriate for your country of use. Slide the switch on the
back of the trainer to select the appropriate input making sure the correct
fuse is installed. This trainer consumes less than 50 W.
Fuse Replacement
Indoor Use
Ground the Instrument
To minimize shock hazard, the instrument case must be connected to an
electrical safety ground. This instrument, and the chassis, is grounded
through the ground conductor of the supplied, three-conductor AC power
cable per IEC 60320-1. Do not alter or bypass the ground connection.
Failure to use a properly-grounded approved outlet and the recommended
three-conductor AC line power cable may result in injury or death. Unless
otherwise stated, a ground connection on the instrument’s front is for a
reference of potential only and is not to be used as a safety ground.
This instrument is intended to be used in an indoor pollution degree 2 environment. The
operating temperature range is 0 °C to 40 °C and the operating humidity range is up to
80% relative humidity with no condensation allowed. Measurement results may be outside
specifications if the instrument is used in non-office-type environments. Such environments
may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical
shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields.
Set the power switch to the OFF or O position and disconnect the power
cord before inspecting or replacing the fuse. Open the black fuse holder
(located at the back of the trainer below the input power receptacle) with
a small, flat-head screwdriver. Remove the old fuse, replace it with a new
one, and reinstall the fuse holder. Use only 5 x 20 mm type T 1.5/250 V or
0.75A/250 V Fuses.
CAUTION
CAUTION
WARNING