PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
Rev. 6-501
56
Table B -- ANALYZER CONFIGURATION SETTINGS (RANGES/CHOICES and DEFAULTS -- continued)
Displayed Screen Title
Entry Range or Choices (where applicable)
Factory Default
Your Setting
RELAY Configuration Settings (continued)
Settings Common To ALARM and CONTROL Relays (continued):
SET TRANSFER?
DE-ENERGIZED or ENERGIZED
All Relays: DE-ENERGIZED
OFF DELAY?
0-300 seconds
0 seconds
ON DELAY?
0-300 seconds
0 seconds
Settings for only ALARM Relays:
LOW ALARM?
pH:
-2.00 to +14.00 pH
ORP: -2100 to +2100 mV
TEMP:-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
pH:
0.00 pH
ORP:
0 mV
TEMP: 0.0
°
C or 32.0
°
F
HIGH ALARM?
pH:
-2.00 to +14.00 pH
ORP: -2100 to +2100 mV
TEMP:-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
pH:
14.00 pH
ORP:
+2000 mV
TEMP: 200.0
°
C or 392.0
°
F
LOW DEADBAND?
pH:
0-10% of range
ORP:
0-10% of range
TEMP: 0-10% of range
pH:
0.00 pH
ORP:
0 mV
TEMP: 0.0
°
C/
°
F
HIGH DEADBAND?
pH:
0-10% of range
ORP:
0-10% of range
TEMP: 0-10% of range
pH:
0.00 pH
ORP:
0 mV
Temp: 0.0
°
C/
°
F
Settings for only CONTROL Relays:
PHASE?
HIGH or LOW
All Relays: HIGH
SET SETPOINT?
pH:
-2.00 to +14.00 pH
ORP: -2100 to +2100 mV
TEMP:-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
pH:
14.00 pH
ORP:
+2000 mV
TEMP: 200.0
°
C or 392.0
°
F
DEADBAND?
pH:
0-10% of range
ORP:
0-10% of range
TEMP: 0-10% of range
pH:
0.00 pH
ORP:
0 mV
TEMP: 0.0
°
C/
°
F
OVERFEED TIMER?
0-999.9 minutes
0 minutes
Settings for only TIMER Relays:
INTERVAL?
0-999.9 minutes
5 minutes
DURATION?
0-999 seconds
5 seconds
OFF DELAY?
0-999 seconds
1 second
PASSCODE Configuration Setting
SET PASSCODE?
DISABLED or ENABLED
DISABLED
TEST/MAINT Simulation Function Settings
SELECT SIM?
SENSOR or TEMPERATURE
SENSOR
SIM SENSOR?
pH:
-2.00 to +14.00 pH
ORP: -2100 to +2100 mV
TEMP:-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
Present measured value of
selected parameter (pH, ORP or
temperature)