4.3 Set
°
C OR
°
F
(temperature display
format)
•
NTC300:
Configures analyzer for use with an NTC
300 ohm thermistor temperature element (used in all
GLI Differential pH and ORP sensors -- except
Model 6006P4-2000 pure water pH sensor systems
which use a Pt 1000 RTD).
•
PT1000:
Configures analyzer for use with a
Pt 1000 RTD temperature element.
•
PT100:
Configures analyzer for use with a
Pt 100 RTD temperature element.
•
MANUAL:
Configures analyzer for fixed manual
temperature compensation -- for pH measurement
only -- when not using a temperature element.
4.
With desired choice displayed, press
ENTER key
t o
enter this selection. When “MANUAL” is selected, you
must set the specific manual temperature compensation
value:
A.
With the
screen displayed, use
Ø
Ø
key
to select the “SET MANUAL” line.
B.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to the desired fixed temperature, and
press
ENTER key
to enter it.
The MEASURE screen can be set to display temperature
values in
°
C or
°
F. In either case, the temperature display
format is always “XX.X.”
1.
With the
sub-submenu screen displayed,
press
ESC key
twice to display