PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 6-501
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
55
4.7 Configuration Setting
Summary
Table B lists all configuration settings and their entry ranges/
choices and factory defaults, categorized by basic functions.
Table B -- ANALYZER CONFIGURATION SETTINGS (RANGES/CHOICES and DEFAULTS)
Displayed Screen Title
Entry Range or Choices (where applicable)
Factory Default
Your Setting
LANGUAGE Configuration Setting
LANGUAGE?
ENGLISH, FRENCH, GERMAN, SPANISH, etc.
ENGLISH
SENSOR Configuration Settings
SELECT SENSOR?
DIFF pH, COMBINATION pH, DIFF ORP or
COMB ORP
DIFF pH
DISPLAY FORMAT?
XX.XX pH or XX.X pH
XX.XX pH
SELECT BUFFER?
“4, 7, 10” or DIN 19267
“4, 7, 10”
PURE H2O COMP
SELECT TYPE?
NONE, AMMONIA MORPHOLINE or USER
DEFINED
NONE
SET ISO POINT?
2.00-10.00 pH
7.00 pH
SET FILTER?
0-60 seconds
0 seconds
PULSE SUPPRESS?
OFF or ON
OFF
ENTER NOTE?
Enter up to eight characters to replace PH
PH
TEMP ELE:SELECT TYPE?
NTC300, PT1000, PT100 or MANUAL
NTC300
TEMP ELE:SET MANUAL? 0.0-100.0
°
C
25.0
°
C
TEMPERATURE Display Configuration Setting
CONFIGURE:
°
C OR
°
F?
°
C or
°
F
°
C
OUTPUT Configuration Settings
SET PARAMETER?
SENSOR or TEMPERATURE
Output 1: SENSOR
Output 2: TEMPERATURE
SET 4mA VALUE?
pH:
-2.00 to +14.00 pH
ORP: -2100 to +2100 mV
TEMP:-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
pH:
0.00 pH
ORP:
0 mV
TEMP: 0.0
°
C or 32.0
°
F
SET 20mA VALUE?
pH:
-2.00 to +14.00 pH
ORP: -2100 to +2100 mV
TEMP:-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
pH:
14.00 pH
ORP:
+2100 mV
TEMP: 200.0
°
C or 392.0
°
F
SET TRANSFER?
0-20 mA or 4-20 mA
All Outputs: 12 mA
SET FILTER?
0-60 seconds
All Outputs: 0 seconds
SCALE 0mA/4mA?
0 mA or 4 mA
All Outputs: 4 mA
RELAY Configuration Settings
Settings Common To ALARM and CONTROL Relays:
SET PARAMETER?
SENSOR or TEMPERATURE
Relay A: SENSOR
Relay B: TEMPERATURE
SET FUNCTION?
ALARM, CONTROL, STATUS or TIMER
All Relays: ALARM
(Table B continued on next page.)