PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 6-501
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
39
Select
DISPLAY FORMAT
SELECT BUFFER Set
for pH Calibration
☞
When using the analyzer to measure ORP, this function
is not provided.
(The ORP display format is fixed to show
mV values as only whole numbers.) For pH measurement,
choose the MEASURE screen display format to be XX.XX
or XX.X. This format setting has no effect on edit/selection
screens, which always show pH values in a XX.XX format.
1.
With the
',63/$< )250$7
6(/(&7 %8))(5
385( +2 &203
6(7 ,62 32,17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use
Ø
Ø
key
to select the “DISPLAY FORMAT” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices (XX.XX or XX.X). With the desired choice dis-
played, press
ENTER key
to enter this selection.
When using the analyzer to measure ORP, this function
is not provided.
For pH, configure the analyzer to use one
of these buffer sets for pH calibration:
•
4.00, 7.00, and 10.00
•
DIN 19267 standard (1.09, 4.65, 6.79, 9.23, and 12.75)
NOTE:
When using buffers that are not included in either of
the analyzer buffer sets, disregard selecting the
buffer set. In this case, use only the “1 (or) 2 POINT
SAMPLE” method for calibration.
The analyzer automatically recognizes pH values from the
selected buffer set and uses its associated built-in table of
pH-versus-temperature values to improve measurement ac-
curacy. To select a buffer set:
1.
With the
',63/$< )250$7
6(/(&7 %8))(5
385( +2 &203
6(7 ,62 32,17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use
Ø
Ø
key
to select the “SELECT BUFFER” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices (4, 7, 10 or DIN 19267). With the desired choice
displayed, press
ENTER key
to enter this selection.