![Gima RESCUE 230 User Manual Download Page 22](http://html1.mh-extra.com/html/gima/rescue-230/rescue-230_user-manual_2228085022.webp)
22
To replace the paper, push the green botton and open the printer cover, insert a new paper roll
with the thermosensible side up and then close the cover. Push the
FEED
key until the paper
comes out straight.
Opening the printer cover
Battery charge
When the ‘
BAT LOW
’ message is displayed please connect the AC charger.
The battery charge light should go on. When the batteries are fully charged the light will go off.
To see the battery charge status, switch on the RESCUE 230 with at least one pad disconnected.
The battery charge status will be displayed. If it shows less than 75% than please connect the AC
charger.
Date/Time setup
To set-up the real time clock, switch on the
RESCUE 230
with at least one pad disconnected.
On the screen will be displayed the battery status, the date and time.
Please use only the original charger.
Do not leave the AC charger connected more than 5 hours. If after this
time the charging light does not go off, please contact the service center.