background image

Cleaning and disinfection

The best way to clean the thermometer tip is by applying a disinfectant (e.g. 70% medical alcohol) with 

a damp cloth. It shall be disinfected before each use. The Waterproof thermometers can be immersed in 

liquid or lukewarm water for through cleaning and disinfection.
Summary of use specification

This usability engineering process assesses and mitigated risks caused by usability problems associated 

with correct use and use errors, it shows the digital thermometer is complied with and acceptance criteria 

documented in the usability validation plan have been met, then the residual risk as defined in ISO14971, 

associated with usability of a medical device are acceptable.
Safety precautions

•  Do not allow the device to come into contact with hot water.

•  Do not expose to high temperatures or direct sunlight. 

•  Do not drop the thermometer. It is neither shock-proof nor impact-resistant.

•  Do not modify this device without the authorization of the manufacturer.

•  Do not bend or open the device (except the battery compartment).

•  Do not clean with thinners, petrol or benzene. Only clean with disinfectant.

•  Do not immerse the Waterproof thermometers under water 15cm for longer than 30 minutes.

•  The thermometer contains small parts (battery, battery compartment) which can be swallowed by chil-

dren. For this reason, do not leave the thermometer unattended in the hands of children.

•  Avoid bending the thermometer tip which contact patient with stainless steel cover

•  If the ambient temperature is over 40℃, dip the thermometer tip in cold water for approx. 5 to 10 seconds 

prior to measuring the temperature.

•  Persistent fever, in particular in children, has to be treated by a doctor- please get in touch with your 

doctor!

•  Do not use near strong electromagnetic fields, i.e. keep it away from any radio systems and mobile

phones.

Battery replacement

The battery is empty and needs replacing when the 

 or 

 battery symbol appears on the right of the 

LCD. Remove the battery cover and remove the battery by toothpick, replace it with a battery (preferably 

non-mercury) of the same type. Please note: The “+ “ sign up and “ - “ sign down.

We advise you to remove the batteries if the device is not going to be used for a longer period of time.
Technical data

Type: maximum thermometer

Measurement range: (32.0~ 42.9)°C 

Measurement accuracy:

+/- 0.1°C

(35.5°C~42.0°C)

+/- 0.2°C

(32.0°C~35.5°C, 42.0°C~42.9°C)

Storage/transportation temperature: (-25~55)°C, ≤95%RH

Ambient temperature during use: (5~40)°C, ≤80%RH

Min Scale: 0.1°C

Atmospheric pressure 700~1060 hPa 

Mode of operation of the clinical thermometer: direct mode

Transient response time: 12s. 

Battery type: alkaline battery, type LR41, 1.5V, service life minimum 100 hours under continuous operation. 

Weight: Approx. 10g

Shelf life: 3 years
Explanation of symbols

Legal requirements and guidelines

This product complies with the European Directive for Medical Device 93/42/EEC and carries the CE mark, 

the device also complies with the specifications of below standard for:

ISO 80601-2-56-2017AMD.1:2018

EN 60601-1

EN 60601-1-11

EN 60601-1-2

The CE Marking confirms that this is a medical device with a measuring function in the sense of the medical 

device Act which has undergone a conformity assessment procedure. A Notified body confirms that this 

product fulfills all the appropriate statutory regulations
Calibration check

This thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to 

the operation instruction, periodic re-adjustment is not required.

The calibration check has to be carried out immediately, if there are indications that the product does not 

keep the defined error limits or the calibration properties could have been affected by an intervention or 

by any other means. Please also observe any national statutory regulations. The calibration check can be 

carried out by the competent authorities or by authorised service providers. A test instruction for calibration 

check can be provided to the relevant authorities and authorised services providers on request.
ELECTROMAGNETIC COMPATIBILITY INFORMATION

This device is suitable for home healthcare environment and professional healthcare facility environment

WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it 

could result in improper operation. If such use is necessary, this equipment and the other equipment should 

be observed to verify that they are operating normally.

The essential performance is the digital thermometer can offer the temperature measurement.

Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromag-

netic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially 

unsafe situation. Recommendation is to keep a minimum distance of 30cm. Verify correct operation of the 

device in case the distance is shorter.

ENGLISH 

DIGITAL THERMOMETER

Note: The exterior of each model has a little difference.

Congratulations on your purchase of this product. Please read the instructions carefully before using the 

thermometer for the first time, and keep these in a safe place. This product is intended for the measurement 

of human body temperature. this product is for home and hospital use, operator shall be at least 11 years 

old and patient can be operator.
Operating Instructions

Before using, please disinfect the probe at first. To switch on, press the ON/OFF button next to the display; 

a short beep will sound, indicating that the thermometer is operational. At the same time the thermometer 

runs a self-check test, during which all the digital segments appear on the LCD. When the letters “Lo” and 

a flashing “°C” display, the thermometer is now ready for use. If the ambient temperature is below 32°C, 

then “Lo°C” will appear on the LCD and if it is more than 42.9°C, then “Hi°C” will appear on the LCD. 

During the reading, the current temperature is displayed continuously and the “°C” symbol flashes. The 

measurement is completed when a constant temperature value has been reached. The temperature value 

is considered constant when the temperature rises less than 0.1°C within 16 seconds. As soon as the 

constant temperature value is reached, a beep will sound 10 times, and the “°C” symbol will stop flashing. 

The highest temperature measured appears on the LCD. However, please note that this thermometer is a 

maximum thermometer, i.e. the displayed temperature can increase slightly if measurement continues after 

the beep. This is particularly the case with axillary measurements, should a temperature value be recorded 

which approximates the core body temperature. In this instance please note the description under “Meth-

ods of measuring temperature”. When the measurement is completed, please switch the thermometer off 

by pressing the ON/OFF button. After the temperature has been displayed, the thermometer will shut off 

automatically after 10 minutes.
Memory function

Switch the thermometer on, a short beep will sound. At the same time the thermometer runs a self-check 

test, during which all the digital segments appear on the LCD. After that the last measured value with “°C” 

will appear automatically on the LCD for about 2 seconds. The reading is only over-written when a new 

temperature value is recorded.
Methods of measuring temperature

It is important to remember that the body temperature reading depends on the site where it is measured. 

For this reason, the measurement site must always be specified in order to ensure that a correct temper-

ature reading is recorded.

In the rectum (rectal) 

This is the most accurate method from a medical point of view, because it comes closest to the core body 

temperature. The thermometer tip is inserted carefully into the rectum for a maximum of 2cm.

The usual measuring time is approximately 20 to 70 seconds.

Under the arm (axillary) 

Placing the thermometer in the armpit provides a measurement of surface temperature that can fluctuate by 

around 0.5°C to 1.5°C from rectal temperature readings in adults. The usual measuring time for this method 

is approximately 40 to 100 seconds. It should be noted, however, that an exact reading cannot be obtained 

if, for example, the armpits have been allowed to cool. If this

is the case, we recommend extending the measuring time by around 5 minutes in order to obtain the most 

precise possible reading that corresponds as closely as possible to the core body temperature.

In the mouth (oral) 

There are different heat zones in the mouth. As a general rule, the oral temperature is 0.3°C to 0.8°C lower 

than the rectal temperature. To ensure that reading is as accurate as possible, place the thermometer tip 

to the left or right of the root of the tongue. The thermometer tip must have constant contact with the tissue 

during the reading and be placed under the tongue in one of the two heat pockets at the back, keep the 

mouth closed during the reading and breathe evenly through the nose. Do not eat or drink anything before 

the measurement. The usual measuring time is approximately 30 to 80 seconds.

Note: We strongly recommend the rectal method as the most accurate method for identifying the basal 

temperature, and advise you to extend the measuring time by 3 minutes after the beep.

Guidance and manufacture’s declaration – electromagnetic emissions
The device is suitable for use in the specified electromagnetic environment and it has meets the follow-

ing standard’s emission requirements
Phenomenon

Profession healthcare

facility environment

Home healthcare environment

Home healthcare

environment

CISPR 11, Group 1,

Class A or B

CISPR 11, Group 1, Class B

Harmonic distortion

IEC 61000-3-2, Class A 

or not applicable

N/A

Voltage fluctuations

and flicker

IEC 61000-3-3 

or not applicable

N/A

 or 

Battery is empty

WEEE disposal

Lo°C

Temperature under 32°C

HI°C

Temperature over 42.9°C

Stand by

Type BF applied part
Caution: read instructions 

(warnings) carefully

Keep away from sunlight

Keep in a cool, dry place

Consult instructions for use

Manufacturer

Date of manufacture

Medical Device complies with 

Directive 93/42/EEC 
Product code

Lot number

IP22

Covering Protection rate 

Temperature limit

Atmospheric pressure limit

Humidity limit

Direct current

Summary of Contents for m25562

Page 1: ...ροφύσιο θερμομέτρου Sensore del termometro Thermometric sensor Capteur thermométrique Sensor termométrico Sensor termométrico Αισθητήρας θερμομέτρου Pulsante ON OFF ON OFF Button Bouton ON OFF Botón ON OFF inicio detener Botão ON OFF Πλήκτρο ON OFF Scompartimento della batteria Battery compartment Compartiment à piles Compartimiento de las pilas Compartimento da pilha Κάλυμμα μπαταρίας ...

Page 2: ...ations on your purchase of this product Please read the instructions carefully before using the thermometer for the first time and keep these in a safe place This product is intended for the measurement of human body temperature this product is for home and hospital use operator shall be at least 11 years old and patient can be operator Operating Instructions Before using please disinfect the prob...

Page 3: ...50 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation 217Hz 2 0 3 28 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation 217Hz 0 2 0 3 9 Guidance and manufacture s declaration electromagnetic immunity The device is suitable for use in the specified electromagnetic environment and it has meets the follow ing immunity test levels Higher immunity levels may cause the de...

Page 4: ...25562 M25562 M Rev 0 06 21 IP22 0476 Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com Made in China 95 1060hPa 700hPa 55 C 25 C 0 ...