Disposal: The product must not be disposed of along with other domestic waste. The users must 

dispose of this equipment by bringing it to a specific recycling point for electric and electronic equip-

ment. 

GIMA WARRANTY TERMS

The Gima 12-month standard B2B warranty applies.

Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare locations where med-

ical equipment and/or systems are used. When they are used in close proximity to medical equipment 

and/or systems, the medical equipment and/or systems’ basic safety and essential performance may be 

affected. This device has been tested with the immunity test level in the below table and meet the related 

requirements of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum distance 

between RF wireless communications equipment and this device as recommended below.

Test 

frequency

(MHz)

Band 

(MHz)

Service

Modulation

Maximum

power W)

Distance

(m)

Immunity

test level

(V/m)

385

380-390

TETRA400

Pulse modulation 

18Hz

1,8

0,3

27

450

430-470

GMRS 460

FRS 460

FM ± 5 kHz

deviation

1 kHz sine

2

0,3

28

710

745

780

704-787

LTE Band

13, 17

Pulse modulation

217Hz

0,2

0,3

9

810

870

930

800-960

GSM 800/900

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5

Pulse modulation

18Hz

2

0,3

28

1720

1845

1970

1700-1990

GSM 1800; 

CDMA 1900;

GSM 1900; 

DECT; 

LTE Band 1, 3, 

4, 25; UMTS

Pulse modulation

217Hz

2

0,3

28

2450

2400-2570

Bluetooth, 

WLAN, 802.11 

b/g/n, RFID 

2450, 

LTE Band 7

Pulse modulation

217Hz

2

0,3

28

5240

5500

5785

5100-5800

WLAN 802.11 

a/n

Pulse modulation

217Hz

0,2

0,3

9

Guidance and manufacture’s declaration – electromagnetic immunity
The device is suitable for use in the specified electromagnetic environment and it has meets the follow-

ing immunity test levels. Higher immunity levels may cause the device’s essential performance lost or 

degraded.
Phenomenon

Basic EMC standard

or test method

Professional 

healthcare facility 

environment

Home healthcare

facility environment

Electrostatic discharge

IEC 61000-4-2

+/- 8 kV contact

+/- 2 kV, +/- 4 kV, +/- 8 kV, +/- 15 kV air

Radiated RF EM fields

IEC 61000-4-3

3V/m

80MHz-2.7GHz

80%AM at 1kHz or 2Hz

10V/m

80MHz-2.7GHz

80%AM at 1kHz or 2Hz

1kHz or 2Hz can be specified by the manufacturer

Proximity fields from RF 

wireless communica-

tions equipment

IEC 61000-4-3

See  the  RF  wireless  communication  equipment 

table in “Recommended minimum separation dis-

tances

Rated power 

frequency

magnetic fields

IEC 61000-4-8

30A/m; 50 Hz or 60Hz

Electric fast

transients bursts

IEC 61000-4-4

N/A

For input a.c. power port

d.c. power lines or signal input/output lines whose length exceeding 3m

Surges

IEC 61000-4-5

N/A

Conducted

disturbances

induced by

RF fields

IEC 61000-4-6

N/A

For 1. input a.c. power port;

2. all d.c. power ports connected permanently to cables >3m

3. all patient-coupled cables

4. SIP/SOP whose maximum cable length ≥ 3m

Voltage dips

IEC 61000-4-11

N/A

Voltage interruptions

IEC 61000-4-11

N/A

UT: rated voltage(s); E.g. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at 60Hz

Summary of Contents for m25562

Page 1: ...ροφύσιο θερμομέτρου Sensore del termometro Thermometric sensor Capteur thermométrique Sensor termométrico Sensor termométrico Αισθητήρας θερμομέτρου Pulsante ON OFF ON OFF Button Bouton ON OFF Botón ON OFF inicio detener Botão ON OFF Πλήκτρο ON OFF Scompartimento della batteria Battery compartment Compartiment à piles Compartimiento de las pilas Compartimento da pilha Κάλυμμα μπαταρίας ...

Page 2: ...ations on your purchase of this product Please read the instructions carefully before using the thermometer for the first time and keep these in a safe place This product is intended for the measurement of human body temperature this product is for home and hospital use operator shall be at least 11 years old and patient can be operator Operating Instructions Before using please disinfect the prob...

Page 3: ...50 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation 217Hz 2 0 3 28 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation 217Hz 0 2 0 3 9 Guidance and manufacture s declaration electromagnetic immunity The device is suitable for use in the specified electromagnetic environment and it has meets the follow ing immunity test levels Higher immunity levels may cause the de...

Page 4: ...25562 M25562 M Rev 0 06 21 IP22 0476 Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com export gimaitaly com www gimaitaly com Made in China 95 1060hPa 700hPa 55 C 25 C 0 ...

Reviews: