5. TECHNICAL SPECIFICATIONS
5.1 GENERAL
Equipment Description:
H.F. Electrosurgery Equipment for monopolar minor surgery. The unit provides the facilities to fine
preset the output power level and choose the coagulating effect through the appropriate Crest Factor.
Cabinet:
Bench-top, metal housing, with the frontal and rear panel covered by lexan sheet.
Height: 100 mm
Width:
180 mm
Depth:
180 mm
Weight: 2,8 Kg
Equipment Mains Input:
Selectable 115 or 230 V~, 50/60 Hz.
Maximum Current:
0,5 A (230V) or 1A (115V)
IEC socket with in-line fuses, voltage selector and mechanical mains switch.
Displays:
LED Digital Displays provide the indication of the preset output level as follows:
CUT:
0 – 50
BLEND:
0 – 45
COAG:
0 - 40
Output Power Control by two keys
The use of the control keys allows the user to control very accurately the output power since it is possible
to vary the
preset output power in step as low as just 1 watt.
Max Output Power:
CUT:
50 W +/- 10%
BLEND:
45 W +/- 10%
COAG:
40 W +/- 10%
Reference Load Resistance: 400
No Load Max Voltage:
900 Vpp +/-10%
Crest Factor:
CUT:
1.5
BLEND:
1.8
COAG:
2.5
Output Frequency: 600 kHz +/- 10%
Modulation Frequency:
CUT:
NONE
BLEND:
10
kHz
COAG:
10
kHz
Allowed output time:
Duty Cycle:
10 seconds output-25 seconds pause