DIATERMO 106
GIMA SpA
Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones
MA126IGBEa
23
4.2.1
E
QUIPMENT
M
AINS
I
NPUT
M
ODULE AND
V
OLTAGE
S
ELECTOR
The equipment mains input module provides the connection point for the power of the equipment's
internal electronics. The mains input module incorporates the mains input socket, the main input voltage
selector and in-line mains fuses.
WARNING
: Before the unit is powered up the operator should check that the mains input voltage
indicated on the power input module corresponds to the mains voltage supply to which it is to be
connected, and that the correct mains fuse type for the selected mains voltage has been inserted.
4.2.2
P
OWER
O
N
-O
FF
S
WITCH
POWER ON/OFF mechanical switch is used to control power to the equipment. To power the
equipment, press the switch in the direction of the 1 (i.e. right side).
When the equipment is powered the front panel will illuminate. Pressing the switch in the 0 direction
will cut power to the equipment, this operation allows the switch to be used as a emergency stop switch,
in the event of any fault.
4.3 FRONT PANEL
1 Key for output power decrease
6 COAG key
2 Output power display
7 Pen-holder connector
3 Key for output power increase
8 Patient Plate connector
4 CUT 1 key
9 Foot-switch connector
5 BLEND key
4.3.1
U
SER
K
EYS
The keys are split into two groups, the four operative keys, positioned at the left of the unit, and the two
output power control keys under the digital display.
The two output power control keys are used to increase (+) or decrease (-) the output current.
The function of each of the operative keys is as follows:
Pressing this key Cut current, i. e. pure sine current is preset, to be delivered when the foot-switch output
control is activated. Crest factor of the Cut current is 1.5. Cut current preset condition is indicated by
green LED which will illuminate.