18

1.6 Working Environment

Operating temperature: 

5°C ~ 40°C

Operating humidity:    

30%   80%

Atmospheric pressure:  

70 ~ 106kPa

2 Installation of batteries

1. Open the battery cover (see the beside picture).

2. Insert two AAA size batteries directly.

3. Close the battery cover.

- Two new AAA size batteries can be used for more than 500 times 

measurement.

- The battery power status is indicated on the screen.

Do NOT insert batteries with their polarities reversed. Dispose of the used batteries according to applicable 

local regulations.

3 Operation

3.1 Cautions

In order to get a more credible measuring result, you should pay attention to the following advice

- Do NOT engage in vigorous exercise prior to testing.

- Keep you relax while testing.

- Do NOT move and talk, try to keep as still as possible during testing.

3.2 Start to measure ECG

2. Press the 

 (Power/Back) button (Lasting for 2 seconds) to turn on 

the device.

After boot-strap, the device starts to initialize, and displays the following pictures.

2. After initialization, the device will enter into a preparing phase; this 

phase will last 6 seconds (see the beeside picture), during this period 

the  device  will  judge  whether  the  placed  position  is  befitting.  Please 

adjust your hand-hold position to obtain good ECG signal if the detected 

waveform is not smooth.

3. During the preparing phase, the user can use 4 methods to measure ECG:

- Measure through both hands, name as palm measurement.

- Measure through right hand and chest, name as chest measurement.

- Measure through right hand and left leg, name as leg measurement.

- Measure through ECG lead wire, name as lead wire measurement.

Note: when you use one of methods to measure, if you can’t 

obtain enough intensity of ECG signal to analyze, please try to 

use another measuring method.

I Palm measurement

Hold the device with right hand. Make sure that the palm and fingers touch 

the metal electrode I and II reliably.

Press the electrode III close against the center of the left palm. The ECG 

signal detected by Palm measurement is equivalent to Lead I ECG signal.

Improper operating method:

A. Waving your both hands at will.

B. Both hands loose contact with electrodes during measurement.

C. Not palm contacts the electrode.

ENGLISH

Summary of Contents for Cardio-A Palm ECG

Page 1: ...ENZIONE Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual befo...

Page 2: ...amination is suggested for normal examination of your heart condition 10 Connect Disconnect data cable to the monitor gently and carefully do NOT wring or shake the data cable plug with force to inser...

Page 3: ...prises main board LCD electrodes etc 1 4 Features Two input connections for ECG signal Embedded metal electrodes Lead wire connector optional Two output modes for ECG data LCD display Serial output in...

Page 4: ...ion the device will enter into a preparing phase this phase will last 6 seconds see the beeside picture during this period the device will judge whether the placed position is befitting Please adjust...

Page 5: ...ECG signal detected by Leg measurement is equivalent to Lead II ECG signal Improper operating method A You moved your foot B Wrong holding gesture C The electrode III can not contact bare skin comple...

Page 6: ...m is determined at the beginning of the measurement automatically 2 1 2 and 1 3 2 1 The measuring time lasts about 30 seconds Then enter the analysis phase in turn Note Refer to the period of result d...

Page 7: ...and press to measure ECG Refer to 3 2 for details 3 3 2 The function of waveform review 1 On the main menu screen press the key or of Navigation key to select Waveform Review and press to enter the re...

Page 8: ...esults and analysis Do not attempt self diagnosis 2 5 If the detected signal is too weak the beeside description can be seen Please try again 2 6 On the result description screen press the power back...

Page 9: ...he key or to move the cursor for selecting setting option and then press the key or to set numerical value 3 Press the power back button to complete the setting and return to the previous menu 3 3 5 2...

Page 10: ...ls Notes During uploading the keys and power key are effective After upload turn off the device and restart it again to start measurement Keep the monitor and data cable still during uploading data Un...

Page 11: ...tal electrode with medical alcohol when the measuring signal is too weak Please wipe the surface of the unit with clean cloth or let it dry in air Please take out the batteries if the unit will not be...

Page 12: ...etic interference 4 Weak signal Trouble ECG waveform excursion strong irrelevant waveform Solution 1 Clean skin with soap and water and eliminate skin crumb and grease to make the skin wet and greasel...

Page 13: ...rences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or changed The defected products must be returned only to the d...

Page 14: ...dal Contraction Premature Ventricular Contraction or multi focus premature beat and help cure Testee s Symptom Normal heart beat is followed by a premature beat 4 Bigeminy It is a type of PVC in which...

Page 15: ...beat The display of missing beat may be doubted as Missed beats Symptom arrhythmia a long time interval Indication May be the conduction system of heart blocks it often occurs in functional cardiovas...

Reviews: